Move the toggle switch (located on rear of console) to OFF position before
leaving unit unattended.
DANGER:
Do not open unit housing at anytime. Danger from optical radiation
may exist.
may exist.
WARNING:
Be aware that the metal / plastic cannula on the tips may becom hot
during use. Avoid contact of the cannula with any tissue.
during use. Avoid contact of the cannula with any tissue.
`
WARNING:
Do not aim the laser at metallic or reflective surfaces, such as
surgical instruments or dental mirrors. If aimed directily at these
surfaces the laser beam will reflect and create a potential hazard.
surgical instruments or dental mirrors. If aimed directily at these
surfaces the laser beam will reflect and create a potential hazard.
Safety Features
Energy Monitor
The current monitor measures and verifies power output. Power deviations of more
than ± 20% from the selected value will cause the display to show the error
message: “DIODE CALIBRATION”.
The unit will not operate until the system is reset by pressing the “Next” button on the
touchscreen and then going into READY mode. If the error messages persist, please
contact Biolase Service at 1-800-321-6717.
The current monitor measures and verifies power output. Power deviations of more
than ± 20% from the selected value will cause the display to show the error
message: “DIODE CALIBRATION”.
The unit will not operate until the system is reset by pressing the “Next” button on the
touchscreen and then going into READY mode. If the error messages persist, please
contact Biolase Service at 1-800-321-6717.
System Monitor
The system monitors the emergency stop switch, remote key, wireless footswitch
connection, and output power. An error in any one of these will stop the system. The
text display will indicate the type of error. Operation will not resume until the error is
cleared.
The system monitors the emergency stop switch, remote key, wireless footswitch
connection, and output power. An error in any one of these will stop the system. The
text display will indicate the type of error. Operation will not resume until the error is
cleared.
Power Switch
The unit can be switched ON or OFF using a toggle switch at the back panel.
The unit can be switched ON or OFF using a toggle switch at the back panel.
Figure 1: Power Switch
Figure 2: Power Supply with cord