  Move the toggle switch (located on rear of console) to OFF position before 
leaving unit unattended. 
 
DANGER:
  
Do not open unit housing at anytime. Danger from optical radiation 
may exist.
 
 
 
WARNING:
 
Be aware that the metal / plastic cannula on the tips may becom hot 
during use.  Avoid contact of the cannula with any tissue. 
 
 
WARNING:
 
Do not aim the laser at metallic or reflective surfaces, such as 
surgical instruments or dental mirrors.  If aimed directily at these 
surfaces the laser beam will reflect and create a potential hazard. 
 
Safety Features 
Energy Monitor 
The current monitor measures and verifies power output. Power deviations of more 
than ± 20% from the selected value will cause the display to show the error 
message: “DIODE CALIBRATION”. 
The unit will not operate until the system is reset by pressing the “Next” button on the 
touchscreen and then going into READY mode. If the error messages persist, please 
contact Biolase Service at 1-800-321-6717. 
 
System Monitor 
The system monitors the emergency stop switch, remote key, wireless footswitch 
connection, and output power. An error in any one of these will stop the system. The 
text display will indicate the type of error. Operation will not resume until the error is 
cleared. 
 
Power Switch 
The unit can be switched ON or OFF using a toggle switch at the back panel.  
               
 
Figure 1: Power Switch
 
Figure 2: Power Supply with cord