EPIC™10 User Manual 
pg. 13 
5400525 Rev B  
 
 
 
LASER 
WARNING: 
Always ensure that the protective laser eyewear is appropriate for the 
laser wavelength.
 
 
•  Do not look directly into the beam or at specular reflections. 
•  Never direct or point the beam at a person’s eyes. 
•  Always place the system into STANDBY mode (by pressing the Control Button while in 
READY mode) before exchanging Handpieces or disposable tips. 
•  Toggle the ON/OFF switch (located on the rear of the console) to the OFF (O) position 
before leaving unit unattended. 
 
LASER 
WARNING: 
Do not open unit housing at any time. Danger from optical radiation may 
exist. 
 
 
LASER 
WARNING: 
Do not aim the laser at metallic or reflective surfaces, such as surgical 
instruments or dental mirrors. If aimed directly at these surfaces the laser 
beam will reflect and create a potential hazard. 
 
CAUTION: 
Be aware that the metal / plastic cannula on the tips may become hot during 
use. Avoid contact of the cannula with any tissue. 
 
3.3 SAFETY FEATURES 
Energy Monitor 
The energy monitor measures and verifies power output. Power deviations of more than ± 20% 
from the selected value will cause the display to show the error message: “LASER CURRENT 
HIGH/LOW”. 
The laser console will not operate until the system first clears the error and then goes into 
READY mode. If the error message persists, please contact BIOLASE Service at                       
1-800-321-6717. 
System Monitor 
The system monitors the emergency stop switch, remote key, wireless footswitch connection, 
and output power. An error in any one of these will stop the system. The text display will indicate 
the type of error. Operation will not resume until the error is cleared.