7.3 SOFT TISSUE SURGERY AND OTHER DENTAL USE 
Tip Initiation: Parameters and Method 
Most soft tissue surgical procedures require initiation of the fiber tip.  The TIP INITIATION 
screen will appear (in READY mode) if tip initiation is recommended
; while in the TIP 
INITIATION  screen, initiate the tip by following the steps outlined below.  
Tip Diameter (µm)
 
(Preset) Power (W)
 
Mode
 
400 
1.4 
CW 
300 
1.4 
CW 
200 
Tip initiation not required when used for recommended procedures
 
Figure 7.1 
 
●    Review the table in Figure 7.1 to determine whether the tip requires initiation.                           
 
●    Touch the tip to the surface of the initiation block, without 
      activating the laser (Figure 7.2). 
 
 
 Figure 7.2
 
●    Press the footswitch to activate the laser, allowing the tip to 
      sink into the block.  Pull the tip out when the metal cannula 
      touches the block, still firing until just before the tip is out of 
      the block (Figure 7.3).  
 
   
 
Figure 7.3
 
●   Press the footswitch to activate the laser into the air once, you 
     will see a white flash or the tip will glow (Figure 7.4). 
 
Figure 7.4
 
●   Repeat initiation process as needed to ensure the tip is initiated. 
After tip initiation is completed, press the check mark to access the screen for the selected 
procedure. 
                          
                        
      
 
Figure 7.5 
 
CAUTION: 
If the laser console is in “READY” mode, the laser will fire if the footswitch is 
activated.