Biolase Inc. EPIC-1 ユーザーズマニュアル
EPIC™10 User Manual
pg. 37
5400525 Rev B
Type of Sterilizer
Temperature
Min Time
Drying Time
Gravity Displacement
121°C ( 250°F)
30 minutes
15 – 30 minutes
132°C (270°F)
15 minutes
Dynamic-Air-Removal (Pre-Vacuum)
132°C (270°F)
4 minutes
20 - 30 minutes
134°C (273°F)
Figure 8.1
• Once the cycle is completed, remove the tray and let each sterilized item cool and dry.
The handpiece and tips must remain in the sterilization pouches until used in order
to maintain sterility.
to maintain sterility.
Cleaning the Whitening/Contour Handpiece
The Whitening Handpiece is sold with disposable non-sterile protective shields.
The handpiece and clear protective shield are not autoclavable. The clear protective
shields are intended for one-time use only and should never be reused to avoid cross-
contamination.
shields are intended for one-time use only and should never be reused to avoid cross-
contamination.
To clean the Whitening Handpiece, wipe down the handpiece with gauze and isopropyl alcohol.
Always wipe the disposable shield with alcohol prior to use. Dispose of after single use.
Always wipe the disposable shield with alcohol prior to use. Dispose of after single use.
8.3 INSTALLING/REPLACING THE CONSOLE BATTERY PACK
1. To install or replace the battery pack, remove the battery cover on the underside of the
console using the Phillips screwdriver included with the laser system (Figure 8.1).
2. To remove the battery, grip the battery at the top and pull the cable away from the
connector (Figure 8.2). Do not tug or wrench the cable from the connector.
3. To install the battery, insert the connector wire from the battery to the unit, making sure
the red wire is on the left, and gently place the battery into the compartment (Figure 8.2).
4. Replace the battery cover on the bottom of the unit, using a standard Phillips
screwdriver.
5. Connect the power cord of the DC power supply to the unit and plug into a wall outlet.
Before first use, you should fully charge the battery (at least three (3) hours). Once the
battery is charged, unplug the power cord from the wall outlet and the console. The unit
will run on battery power alone. (See Section 4.1)
battery is charged, unplug the power cord from the wall outlet and the console. The unit
will run on battery power alone. (See Section 4.1)
6. Recycle the used Lithium Ion battery as regulated. Do not throw it in a trashbin.