3. Click the OP Band button to set the operating band that will be used by the
Access Point.
•
•
Typically, the 608.5-613.0 OP Band is selected. However, where the Patient-
Net System operates in the same hospital as another system (or systems) that
also use(s) the 608-614 MHz WMTS band, it will be necessary to select one
of the sub-bands. The sub-band should coordinate with the hospital’s spec-
trum policy.
•
Additional sub-bands include: 608.5-609.7, 609.6-610.8, 610.7-611.9, and
611.8-613.0
WARNING: All transceivers must be programmed to the same OP Band as the Access
Points. This is an administrative procedure only. Do not attempt to program
the transceivers with the selected OP Band.
the transceivers with the selected OP Band.
4. Click the Update All Access Points button to update any allocated Access
Points on the OpenNet LAN with the current configuration data from the
Device Master.
Device Master.
WARNING: Selecting Update All Access Points will re-configure all Access Points on the
network and will affect proper system operation. Do not activate this button
on a live system.
on a live system.
5. Click the Exit button to close the OpenNet LAN Configuration screen and