EN/IEC 60601 
Test Level
 ±1 kV line(s) 
to line(s)±2 kV 
line(s) to earth
<5% U
T
(>95% dip in U
T
for 0.5 cycle. 
40% U
T
 (60% dip 
in U
) for 5 
cycles.
70% U
T
 (30% dip 
in U
) for 25 
cycles
5% U
T
 (>95% dip 
in U
T
 ) for 5 sec.
Compliance 
Level
 ±1 kV line(s) 
to line(s)±2 kV 
line(s) to earth
Immunity Test
Surge EN/IEC 
61000-4-5
(only for battery 
charger power supply)
3A/m
3A/m
Power Frequency 
50/60Hz Magnetic 
Field EN/IEC 
61000-4-8 
Voltage dips, 
short 
interruptions 
and voltage 
variations on 
power supply 
input lines 
EN/IEC 
61000-4-11
(only for battery 
charger power supply)
<5% U
T
(>95% dip in U
T
for 0.5 cycle. 
40% U
T
 (60% dip 
in U
) for 5 
cycles.
70% U
T
 (30% dip 
in U
) for 25 
cycles
5% U
T
 (>95% dip 
in U
T
 ) for 5 sec.
Electromagnetic 
Environment – Guidance
NOTE : U
is the a.c. mains voltage prior to application of the test level.
Electrical power quality 
should be that of a typical 
commercial or hospital 
environment. During charging, 
if the the user of the GoSafe 
Lite 7150MHB requires 
continued operation during 
power mains interruptions, it 
is recommended that the 
GoSafe Lite device power 
adapter is powered from an 
uninterrupted power supply 
or a battery.
Mains power quality should be 
that of a typical commercial or 
hospital environment.
Power frequency magnetic 
fields should be that of a 
typical commercial or hospital 
environment.