Lifeline Systems Inc. 7100MHB ユーザーズマニュアル
EN/IEC 60601
Test Level
±1 kV line(s)
to line(s)±2 kV
line(s) to earth
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle.
40% U
40% U
T
(60% dip
in U
T
) for 5
cycles.
70% U
T
(30% dip
in U
T
) for 25
cycles
5% U
5% U
T
(>95% dip
in U
T
) for 5 sec.
Compliance
Level
±1 kV line(s)
to line(s)±2 kV
line(s) to earth
Immunity Test
Surge EN/IEC
61000-4-5
(only for battery
charger power supply)
3A/m
3A/m
Power Frequency
50/60Hz Magnetic
Field EN/IEC
61000-4-8
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
EN/IEC
61000-4-11
(only for battery
charger power supply)
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle.
40% U
40% U
T
(60% dip
in U
T
) for 5
cycles.
70% U
T
(30% dip
in U
T
) for 25
cycles
5% U
5% U
T
(>95% dip
in U
T
) for 5 sec.
Electromagnetic
Environment – Guidance
NOTE : U
T
is the a.c. mains voltage prior to application of the test level.
Electrical power quality
should be that of a typical
commercial or hospital
environment. During charging,
if the the user of the GoSafe
Lite 7150MHB requires
continued operation during
power mains interruptions, it
is recommended that the
GoSafe Lite device power
adapter is powered from an
uninterrupted power supply
or a battery.
Mains power quality should be
that of a typical commercial or
hospital environment.
Power frequency magnetic
fields should be that of a
typical commercial or hospital
environment.