Chapter 8 Troubleshooting   
8.1 Check the equipment and accessories 
A visual inspection should be carried out before each use when the power is disconnected: 
1.
 
Check the outside of the equipment to see if it is clean and the general physical condition. Make 
sure the outer skin is not broken, all parts are in place, and there is no dumping of liquid as well as 
no signs of abuse.   
2.
 C
heck whether all accessories (such as cables, sensors, sensory unit, etc.)    are in good condition. If 
there is any sign of damage, please stop using it.   
8.2 Check cable and charging line 
1. Check whether the power plug and the power cord are damaged, and make sure the plug is fixed 
and not moved in the outer skin. If there is any damage, please contact the maintenance personnel, 
Vita-Course Technologies or your authorized supplier. 
2. Check the user's cable and its corresponding stress relief device in general to confirm that the 
insulating layer is undamaged and that both ends of the connector are firmly connected to prevent 
rotation or other stresses. 
3.    Clamp the pulse wave sensor on the user's finger and open the equipment and gently pull the 
cable line near the ends of the cable to make sure there is no interruption.   
8.3 Troubleshooting 
If you suspect a measurement results, please read the instruction manual and confirm several times 
that the measurement is properly set. 
If you have any problems with the equipment, please call the maintenance staff.
 
Phenomenon 
 
Possible reason 
 
Solution 
 
The equipment is 
power-off. 
Depletion of internal lithium 
battery   
Plug in the power adapter to charge 
the battery. 
Possible damage 
Send the product to the designated 
repair shop or sales shop for repairs 
No reaction 
when pushing 
Work abnormally 
The equipment turns off if you push 
button no less than 10 seconds; it will