The guidance and manufacture’s statement--electromagnetic emission 
Users  or  buyers  of  the  equipment  should  make  sure  that  it  is  used  in  such  following 
electromagnetic environment: 
Emission test 
Conformity 
electromagnetic environment–guidance 
GB4824 RF emission 
1group 
The device uses RF energy only for its internal 
functions. As a result, its RF emission is very low and 
may not cause any interference to nearby electronic 
devices. 
GB4824 
RF emission 
B type 
The equipment is suitable for use in all facilities, 
including  homes  and  public  low  voltage  power 
grids connected directly to homes. 
Gb17625.1 
Harmonic emission 
Inadequacy 
GB17625.2 
Voltage  fluctuation  / 
scintillation emission 
Conformity 
 
The guidance and manufacture’s statement–electromagnetic immunity 
Users  or  buyers  of  the  equipment  should  make  sure  that  it  is  used  in  such  following 
electromagnetic environment: 
Immunity test 
IEC 60601 test level 
Accord with level 
Electromagnetic 
immunity–guidance 
Electrostatic 
discharge
(ESD) 
GB/T 17626.2 
±6 kV contact 
discharge 
±8 kV air discharge 
±6 kV contact 
discharge 
±8 kV
 
air discharge 
The  ground  shall  be  wood, 
concrete  or  tile,  and  if  the 
ground  is  covered  with 
synthetic 
material, 
the 
relative humidity shall be at 
least 30%. 
EFT--electrical 
fast transient 
GB/T 17626.4 
+ 2kV to power line 
+ 1kV to input / output 
line 
±2kV to power line 
The  network  power  shall 
have typical quality used in 
the  commercial  or  hospital 
environment. 
Surge   
GB/T 17626.5 
+ 1 kV differential 
mode voltage 
+ 2 kV common mode 
voltage 
±1 kV differential 
mode voltage 
The  network  power  shall 
have typical quality used in 
the  commercial  or  hospital 
environment.