Vita-Course Technologies Co. Ltd. DISCOVERY ユーザーズマニュアル
According to the type of protection against electric shock: class II internal power supply equipment
According to the degree of protection against electric shock: BF-type equipment
According to the degree of protection against harmful liquid into: ordinary equipment (the closed
equipment that cannot stop the liquid into)
According to the safety degree related to a mixture of flammable anesthetizing gas and air or
oxygen or Nitrous Oxide: not applicable in the place existing flammable anesthetizing gas.
According to the work system: continuous operation of equipment
According to the degree of protection against harmful liquid into: ordinary equipment (the closed
equipment that cannot stop the liquid into)
According to the safety degree related to a mixture of flammable anesthetizing gas and air or
oxygen or Nitrous Oxide: not applicable in the place existing flammable anesthetizing gas.
According to the work system: continuous operation of equipment
1.6 Operating environment
1.Operating temperature:5 ℃~45 ℃;
Relative humidity:10%~95%( non-condensing);
Atmospheric pressure:70.0 kPa~106.0 kPa;
2.Working mode: internal power supply: DC 3.7, V/2200, mAh;
Charging mode: power adapter input voltage: 220 V, 50Hz.
Relative humidity:10%~95%( non-condensing);
Atmospheric pressure:70.0 kPa~106.0 kPa;
2.Working mode: internal power supply: DC 3.7, V/2200, mAh;
Charging mode: power adapter input voltage: 220 V, 50Hz.
1.7 Safety
1. Comply with IEC 60601-1:2012 “Medical electrical equipment; Part 1: General requirements for
safety”
2. Comply with IEC 60601-1-2:2014 “Medical electrical equipment – Part 1-2: General requirements
for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances –
Requirements and tests”
safety”
2. Comply with IEC 60601-1-2:2014 “Medical electrical equipment – Part 1-2: General requirements
for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances –
Requirements and tests”
3. Comply with ISO 80601-2-61 “Particular requirements for basic safety and essential
performance of pulse oximeter equipment”
performance of pulse oximeter equipment”