Error Messages and Troubleshooting
55
54
CAUTION
A blood sample should ONLY be applied to the test strip after the test strip has been inserted 
correctly and the Meter's screen is showing the image of the test strip and a blood drop 
flashing symbol. If the blood drop flashing symbol is NOT showing on the screen, do not 
apply a sample to the test strip. Please re-insert the unused test strip correctly. The screen 
MUST show the image of the test strip accompanied by the blood drop symbol flashing (this 
symbol take 3 seconds to show after correctly inserting the test strip) in order for a sample to 
be applied. Please consult your User Manual and/or contact Customer Service for support on 
how to correctly insert a GE006 Test strip.
Limitations 
- GE006 Blood Glucose Monitoring System is not intended for serum or plasma testing.
- Inaccurate test results may be obtained at altitudes greater than 10,000 feet (3048 meters) above sea 
level.
- Severe dehydration and excessive water loss may cause inaccurately low results.
- High concentrations of  Uric acid 
10.0 mg/dL, L-Dopa 
3 mg/dL, Dopamine 
2 mg/dL, Ascorbic 
acid (Vitamin C) 
5 mg/dL may result in inaccurate test results.
- Not for neonatal use.
- For over-the-counter use. For single-patient use only. For in vitro diagnostic use only.
- Not for screening or diagnosis of diabetes mellitus.
- Not for use on critically ill patients in shock, dehydrated patients or hypo-osmolar patients. 
- Alternative site testing (AST) should only be performed during steady-state times (when glucose is not 
changing rapidly).
- Alternative Site Testing  should not be used to calibrate continuous glucose monitoring systems 
(CGMS).
- Alternative site testing should not be performed if testing for hypoglycemia (Low blood glucose).
- Results from alternative site testing should not be used in insulin dose calculation.