MEDTRUM TECHNOLOGIES Inc. FM006 ユーザーズマニュアル

ページ / 49
 
44 
 
Alarm level
Condition 
Message 
Cause 
Alarm 
Patch Error 
! ! PATCH ERROR 
Insulin delivery stopped, change patch 
now. ESC,OK to clear 
Detect unexpected Patch error. 
Alarm 
Exceeds 
Max TDD 
! ! EXCEEDS MAX TDD 
Exceeds max TDD, no delivery. ESC,OK 
to clear 
Exceeds max TDD. 
Alarm 
Exceeds 
Max 1 HR 
Delivery 
! ! EXCEEDS MAX 1 HR DELIVERY 
Exceeds max 1 hour MAX, no delivery. 
ESC,OK to clear 
Exceeds max 1 hr delivery. 
Alarm 
Pump Base 
Error 
! ! PUMP BASE ERROR 
Remove patch pump, call customer 
service. ESC,OK to clear 
Pump Base error. 
Alarm 
Empty 
Reservoir 
! ! EMPTY RESERVOIR 
No delivery, change patch now. ESC,OK 
to clear 
There is no insulin in the 
reservoir. Change Patch now. 
Alarm 
PDM Error 
! ! PDM ERROR 
Remove device now, call customer 
support. ESC,OK to clear 
PDM error. 
Alarm 
Replace 
Battery 
! ! REPLACE BATTERY 
Replace the battery. Remove battery to 
silence the alarm. ESC,OK to clear 
The battery is dead. 
Alert 
Failed BATT 
Test 
! FAILED BATT TEST 
Insert new battery. Remove battery to 
silence the alarm. ESC,OK to clear 
The battery does not have 
enough voltage.