Figure 11:  Front location incorrect placement 
 
5.5
 
REPLACING THE PATCH  
5.5.1
  REMOVE THE SENSOR FROM THE PATCH 
1.
  Press the blue upper snap-in clip.  
2.
  Tilt and remove the Sensor from the Patch while pressing the clip. 
3.
  Place the Sensor in the Charger.  Whenever the Sensor is not on the body, it should be 
kept in the Charger. 
 
5.5.2
  REMOVE THE PATCH FROM THE BODY 
1.
  Hold the skin with one hand while using the other to gently peel-off the Patch. You may 
use a wet cloth to assist you in the process 
2.
  Discard the Patch immediately.  Do not reuse. 
3.
  If required, replace the Patch with a new one according to the instructions in Sections 
5.2 to 5.4.  The Patch should be replaced at least once every five days.  Make sure to