ZOLL Medical I Ltd. UCOR30 ユーザーズマニュアル

ページ / 29
 
µCor 3.0 Monitoring System 
Instructions for Use
 
 
 
  
                    
Page 23 of 29 
 
10
  TECHNICAL SPECIFICATIONS 
10.1
  ACCURACIES 
Parameter 
Measurement Accuracy 
Thoracic 
Impedance* 
Error of less than 50cc fluid at a confidence level of 90% 
HR 
Accuracy higher than ±2 bpm or ±2%, the higher of the two 
RR 
Maximum error of ±2 bpm 10% , the higher of the two 
Activity 
Activity classification accuracy > 90%**  
Posture 
Posture classification accuracy > 90% 
 
*
 
As measured in experiments conducted in the laboratory 
** resting vs. active > 90% & walking (vs. non-walking) > 90% 
 
Heart Rate Calculation  
The µCor 3.0 system algorithm detects the peak of each R-wave and calculates the time interval 
between successive R peaks known as the RR interval.  Heart Rate is calculated based on a 60 
second moving average time window. For each window where valid ECG was detected, Heart 
Rate is computed in bpm as 60 divided by the average RR interval.  
 
10.2
  ELECTRICAL (POWER) REQUIREMENTS 
Sensor 
Max Power Consumption 
0.85W 
Nominal Power Consumption  
0.02W 
Max Rating 
4.2V, 0.2A 
Nominal Rating 
3.7V, 0.005A 
Battery Type  
Li-Pol, 3.7V, 370mAh