Innokas Yhtyma Oy VC150 ユーザーズマニュアル
KO00065K
VC150 Vital Signs Monitor
7-9
GE TruSignal SpO
2
: SpO
2
procedure
SpO
2
measurement site
For documentation purposes, you can select the measurement site for SpO
2
in
Monitor Setup > SpO
2
before you start measuring SpO
2
. A pop-up screen with
the following images will appear: Finger, Nose, Toe, Earlobe, Other or None. It
does not matter whether the site is on the left or right side of the patient. After
the site is selected, it will be displayed on the home screen in the SpO
does not matter whether the site is on the left or right side of the patient. After
the site is selected, it will be displayed on the home screen in the SpO
2
parameter window. The selection can be changed between the measurements.
Admitting a new patient will reset the selection of the measurement site.
Admitting a new patient will reset the selection of the measurement site.
SpO
2
procedure
1. Check the label on the monitor to determine which SpO
2
technology the
monitor is using. To assure optimal performance, use only accessories that
are intended for that technology. If you cannot read the label, ask the nurse
manager or service which SpO
are intended for that technology. If you cannot read the label, ask the nurse
manager or service which SpO
2
technology is used.
2. Select a sensor that is appropriate for the patient, the clinical situation and
for the SpO
2
technology used. Do not use an adult sensor on a neonatal/
pediatric patient and vice versa.
WARNING
Do not use a sensor, cables, or connectors that appear damaged or
with exposed electrical contacts.
with exposed electrical contacts.
Never repair a damaged sensor or cable; never use a sensor or cable
repaired by others.
repaired by others.
3. Ensure the VC150 is connected to power or the battery is fully charged.
4. Ensure the VC150 is powered on. The monitor has to display clinical mode
and SpO
2
parameter on screen.
5. Connect an interface cable to the monitoring system sensor port.