Setup
4.2 Testing the Basic Functions
A functional test should be performed after initial installation, following each step
below:
below:
1. Remove TOCO-transducer, while the US-transducer is charging:
•
charge bay LED and battery indicator LEDs turn off
•
LED in TOCO-transducer starts fl ashing
•
no further action (no signal transmission, because the US-transducer is not
running)
running)
2. Dock TOCO-transducer for charging again:
•
charge bay LED and battery indicator LEDs turn on
3. Remove US-transducer (US):
•
charge bay LED turns off
•
LED in US-transducer starts fl ashing
•
US-transmission indicator LED turns green
•
US battery indicator LEDs indicate charge status
•
US channel becomes active: US is indicated on fetal monitor
4. Simulate audio signal:
•
Stroke bottom face of transducer at approximately 2 strokes per second to
simulate a fetal heart signal.
simulate a fetal heart signal.
•
fetal monitor shows heart rate, after a short delay.
5. Remove the TOCO-transducer:
•
charge bay LED turns off
•
LED in TOCO-transducer starts fl ashing
•
TOCO-transmission indicator LED turns green
•
TOCO battery indicator LED's indicate charge status
•
TOCO channel becomes active: TOCO is indicated on fetal monitor
Note :
The TOCO transducer must be in operating range of the US
transducer (Maximum range 30cm).
6. Press gently on the sensor area of the transducer:
•
fetal monitor shows adequate TOCO values
7. Dock US-transducer for charging
•
US and TOCO-transmission indicator LEDs turn Amber
•
charge bay LED turns on
•
US battery bay LEDs indicate charge status
•
Fetal monitor returns to normal operating mode with wired transducers.