Setup
4.2  Testing the Basic Functions
A functional test should be performed after initial installation, following each step 
below:
1.  Remove TOCO-transducer, while the US-transducer is charging:
• 
charge bay LED and battery indicator LEDs turn off
• 
LED in TOCO-transducer starts fl ashing
• 
no further action (no signal transmission, because the US-transducer is not 
running)
2. Dock TOCO-transducer for charging again: 
• 
charge bay LED and battery indicator LEDs turn on
3. Remove US-transducer (US):
• 
charge bay LED turns off
• 
LED in US-transducer starts fl ashing
• 
US-transmission indicator LED turns green
• 
US battery indicator LEDs indicate charge status 
• 
US channel becomes active: US is indicated on fetal monitor
4. Simulate audio signal:
• 
Stroke bottom face of transducer at approximately 2 strokes per second to 
simulate a fetal heart signal.
• 
fetal monitor shows heart rate, after a short delay.
5. Remove the TOCO-transducer:
• 
charge bay LED turns off
• 
LED in TOCO-transducer starts fl ashing
• 
TOCO-transmission indicator LED turns green
• 
TOCO battery indicator LED's indicate charge status
• 
TOCO channel becomes active: TOCO is indicated on fetal monitor
Note :  
The TOCO transducer must be in operating range of the US  
   
transducer (Maximum range 30cm).
6. Press gently on the sensor area of the transducer: 
• 
fetal monitor shows adequate TOCO values
7. Dock US-transducer for charging
• 
US and TOCO-transmission indicator LEDs turn Amber
• 
charge bay LED turns on
• 
US battery bay LEDs indicate charge status
• 
Fetal monitor returns to normal operating mode with wired transducers.