iPRECIO
®
 SMP/UCD-300 Quick guide ver.1.0.0.j 
 
 
Pump detection 
 
 
1.  Place pumps by recommended position with data com unit UCD-300. 
2.  Click  Detect  button.  During  Communication  progress  bar  is  shown,  the 
communication enable pump will be detected and assign individual animal ID. 
 
 
 
If there is no response from pump, please check the followings: 
 
It  may  take  long  term  to  detect  pump  due  to  radio  wave  back 
ground environment. 
 
Verify  PC  and  UCD-300  connect  properly.  For detail, please refer 
“TCP/
IP connection between data com unit and control PC” and 
“Network Setting”.
 
 
Verify pump activation. For detail, please refer “Catheter connection, 
plomb admin liquid and activation”.
 
 
3.  After all pump detected, then click Next button. 
 
 
KVO setting, start and complete 
 
KVO is abbreviate of Keep Vein Open, semantics to prevent the catheter edge clog by 
continuous  liquid  dosing  by  low  flow  rate.  Be  normally  administrate  biocompatible 
liquid such as saline as KVO by low flow rate after implant surgery and prevent clog of 
catheter end during animal healing period. After animal comeback form the surgery, 
the KVO liquid in reservoir can be replaced by medicine liquid from the septum. And 
acknowledge user the timing that when the medicine liquid will reach the catheter end 
by pump and appurtenant catheter dead volume calculation. 
 
 
1.  KVO setting window pops up. If KVO is not required, then check skip KVO and click 
Next>> button. 
2.  Click individual column dead volume setting. 
3.  Input the required information for dead volume calculation and click OK button. 