Olympus Winter & Ibe GmbH ESG200 ユーザーズマニュアル
68
Inspection before operation
To test the functionality of the output power settings proceed as follows:
1. Tap into the pane [MONOPOLAR].
Check that the screen changes from the All Screen to the Set Screen, displaying the
settings of the MONOPOLAR output socket.
2. Test the plus buttons by tapping [+].
For values up to 50 the output power level increases in steps of 1 per tap.
For values above 50 the plower level increases in steps of 5 per tap.
3. Test the minus buttons by tapping [−].
For values up to 50 the output power level decreases in steps of 1 per tap.
For values above 50 the plower level decreases in steps of 5 per tap.
4. Touch and hold [+] and then [−] to check that the output power level increases/
decreases fast.
5. Tap [Return] to return to the All Screen.
Testing the screen [Select Menu]
Figure 9.2 Changing from the All Screen to the screen [Select Menu]
To test the functionality of the screen [Select Menu] proceed as follows:
1. Tap [MENU] on the All Screen.
Check that the screen changes from the All Screen to the screen [Select Menu].
2. Tap [Return] to return to the All Screen.
9.3 Inspecting the foot switch assignment
Figure 9.3 Changing from the All Screen to the screen [Assign Foot Switch]
To test the functionality for the foot switch assignment proceed as follows:
1. Tap [FOOT SWITCH] on the All Screen.
DRAFT
–