Electromagnetic compatibility
Guidance and manufacturer’s declaration – electromagnetic immunity
This device is intended for use in the electromagnetic environment specified below. The customer or 
user of this device should assure that it is used in such an environment.
Immunity test IEC 60601 
test level
Compliance 
level
Electromagnetic environment – guidance
Conducted RF 
according to 
IEC 61000-4-6
3 V
eff
150 kHz to 
80 MHz
3 V
Portable and mobile RF communications equipment 
should be used no closer to any part of the Olympus 
CELON ELITE ESG-200, including cables, than the 
recommended separation distance calculated from the 
equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2 √ P
d = 1.2 √ P 80 MHz to 800 MHz
d = 2.3 √ P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the 
transmitter in watts [W] according to the transmitter 
manufacturer and d is the recommended separation 
distance in meters [m].
Field strengths from fixed RF transmitters, as determined 
by an electromagnetic site survey
a
, should be less than the 
compliance level in each frequency range
b
.
Interference may occur in the vicinity of equipment marked 
with the following symbol:
Radiated RF 
according to 
IEC 61000-4-3
3 V/m
80 MHz to 
2.5 GHz
3 V/m
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is 
affected by absorption and reflection from structures, objects and people.
a
 Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) 
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV 
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic 
environment due to fixed RF transmitters, an electromagnetic site survey should be 
considered. If the measured field strength in the location in which the Olympus CELON 
ELITE ESG-200 is used exceeds the applicable RF compliance level above, the Olympus 
CELON ELITE ESG-200 should be observed to verify normal operation. If abnormal 
performance is observed, additional measures may be necessary, such as re-orienting or 
relocating the device.
b
 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances between portable and mobile RF communications 
equipment and this device
This device is intended for use in an electromagnetic environment in which radiated RF disturbances 
are controlled. The customer or user of this device can help prevent electromagnetic interference 
by maintaining a minimum distance between portable and mobile RF communications equipment 
(transmitters) and this device as recommended below, according to the maximum output power of the 
communications equipment.
DRAFT 
– 