Olympus Winter & Ibe GmbH ESG200 ユーザーズマニュアル
94
Electromagnetic compatibility
Guidance and manufacturer’s declaration – electromagnetic immunity
This device is intended for use in the electromagnetic environment specified below. The customer or
This device is intended for use in the electromagnetic environment specified below. The customer or
user of this device should assure that it is used in such an environment.
Immunity test IEC 60601
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic environment – guidance
Conducted RF
according to
IEC 61000-4-6
3 V
eff
150 kHz to
80 MHz
3 V
Portable and mobile RF communications equipment
should be used no closer to any part of the Olympus
CELON ELITE ESG-200, including cables, than the
recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2 √ P
d = 1.2 √ P 80 MHz to 800 MHz
d = 2.3 √ P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts [W] according to the transmitter
manufacturer and d is the recommended separation
distance in meters [m].
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey
a
, should be less than the
compliance level in each frequency range
b
.
Interference may occur in the vicinity of equipment marked
with the following symbol:
Radiated RF
according to
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the Olympus CELON
ELITE ESG-200 is used exceeds the applicable RF compliance level above, the Olympus
CELON ELITE ESG-200 should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or
relocating the device.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances between portable and mobile RF communications
equipment and this device
This device is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or user of this device can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and this device as recommended below, according to the maximum output power of the
communications equipment.
DRAFT
–
2017-01-17