Clarity - INT9 A/D converter 
 Troubleshooting 
30 
If your A/D card is not in the left-hand 
Setup 
Control Modules
 list, add it using the 
Add
 
button and repeat the previous step. 
Note: 
More information on configuration of the INT9 card is 
in chapter 2.4 - Clarity configuration on pg. 9. 
b)  You have a licence purchased for collections from a smaller 
number of instruments: 
Solution: 
Open the 
System Configuration
 dialog from the 
main 
Clarity
 window using the 
System - 
Configuration
 command and check the tab of 
the corresponding instrument – 
Instrument x
Blue symbol of the curve in the tab header 
indicates an Instrument not usable for 
measuring. 
Check your serial number S/N (for example) 
using the 
Help - About
 command from the 
main 
Clarity
 window. 
c)  You are using Clarity Offline or a DEMO version, which does 
not enable the measurement of chromatograms. 
Solution: 
Check whether there is a blue line with the title 
OFFLINE  displayed in the main 
Clarity
 window 
under the symbols of the instruments, or the 
title DEMO in the window header. 
4.5  Data Acquisition - Simulated 
The label “Simulated” is displayed in the Data Acquisition window  
The corresponding instrument only displays 
the simulated curve (from the file 
CHANNX.DTA
), 
or is allocated as what is known as a DEMO 
driver. 
Solution: 
Open the 
System Configuration
 dialog from the 
main 
Clarity
 window using the 
System - 
Configuration
 command and check the tab of 
the corresponding instrument - 
Instrument x
From the list of equipment allocated to the 
instrument, remove the 
Detector x from DEMO 
driver
 field and draw the correct detector of the 
A/D card you are using from the list on the 
left.