Voltcraft DLA-1L 16 ユーザーズマニュアル

ページ / 6
 
OPERATING INSTRUCTIONS
VERSION 03/15
DLA-1L 16 MEASURING ADAPTER
ITEM NO.: 1271059
INTENDED USE
The measuring adapter measures together with a clamp meter or multimeter (clamp meter 
and multimeter are not included in the delivery content) the electrical current and voltage of 
single-phase devices with protective contact or euro plugs. The measuring adapter is inserted 
between the electrical consumer and the socket. The measuring adapter may only be used in 
the range of the overvoltage category CAT III in AC voltage networks with a nominal voltage 
of max. 250 V/AC which are fused with 16 A.
Only use the measuring adapter for the duration of the measurement. A permanent retention 
in the mains cable circuit is not permitted.
The measuring operation is only permitted in dry surroundings. 
For safety and approval purposes (CE), you must not rebuild and/or modify this product. If 
you use the product for purposes other than those described above, the product may be 
damaged. In addition, improper use can cause hazards such as short circuiting, fire, electric 
shock etc. Read the instructions carefully and keep them. Make this product available to third 
parties only together with its operating instructions.
This product complies with the statutory national and European requirements. All company 
names and product names are trademarks of their respective owners. All rights reserved.
DELIVERY CONTENT
•  Measuring adapter with protective contact socket and protective contact plug
•  Operating instructions
SAFETY INSTRUCTIONS
Read the operating instructions carefully and especially observe the safety 
information. If you do not follow the safety instructions and information 
on proper handling in this manual, we assume no liability for any resulting 
personal injury or damage to property. Such cases will invalidate the 
warranty/guarantee.
a) Persons / Product
•  The device is not a toy. Keep it out of the reach of children and pets.
•  Do not leave packaging material lying around carelessly. These may become 
dangerous playing material for children.
•  Protect the product from extreme temperatures, direct sunlight, strong jolts, high 
humidity, moisture, flammable gases, vapours and solvents.
•  Do not place the product under any mechanical stress.
•  If it is no longer possible to operate the product safely, take it out of operation and 
protect it from any accidental use. Safe operation can no longer be guaranteed if 
the product:
 - is visibly damaged, 
 - is no longer working properly, 
 - has been stored for extended periods in poor ambient conditions or 
 - has been subjected to any serious transport-related stresses.
•  Please handle the product carefully. Jolts, impacts or a fall even from a low height 
can damage the product.
•  Also observe the safety and operating instructions of any other devices which are 
connected to the product.
•  The measuring adapter and the plugs may not be disassembled.
•  The voltage and current range specified should not be exceeded.
•  The measuring adapter may only be used for current and voltage measurement.
•  The user is not allowed to put other items than suitable safety measuring cables 
in the sockets
•  Current measurement on the measurement sockets is not possible (an attempt 
would short-circuit the device)! 
•  The user may not short-circuit the contacts of the measuring adapter.
b) Miscellaneous
•  Consult an expert when in doubt about operation, safety or connection of the 
device.
•  Maintenance, modifications and repairs are to be performed exclusively by an 
expert or at a qualified shop.
If you are not sure about the correct connection or use, or if questions arise which are not 
covered by these operating instructions, please do not hesitate to contact our technical 
support or another qualified specialist.
OPERATING ELEMENTS
2
5
3
1
4
1  Protective contact socket
2  Measuring chamber PE
3  Cover for measuring sockets
4  Measuring socket PE 
5  Measuring socket 
6  Measuring socket L
7  Measuring chamber L
8  Protective contact plug
7
6
8
OPERATION
The current is measured with commercial AC current clamps on the insulated 
measuring chambers and the voltage is measured with voltage measuring devices 
on the measuring sockets. Depending on the position of the plug, the current can be 
measured from measuring chamber L or N and, not depending on the position, also 
the leakage current (stray current) can be measured at the measuring chamber PE
Both types of measurement can be carried out at the same time.
Only use the measuring adapter for the duration of the measurement. A permanent 
retention in the mains cable circuit is not permitted.
CAT III Overvoltage category III for measuring in building installation (e.g. outlets or 
sub-distribution). This category also covers all smaller categories (e.g. CAT II for 
measuring electronic devices).
The measuring adapter allows for easy current measurement in closed line systems 
with power plugs. Both main conductors phase L and protective conductor PE are freely 
accessible via the insulated measuring chamber without opening the mains cable at great 
expense. The current via L respectively N can be measured at the measuring chamber 
L depending on the plug position. The measuring chambers are correspondingly labeled 
(L, PE).
The measuring sockets are 
for voltage measurement.
Current clamp is 
not included in the 
delivery content.
Proceed as follows to conduct a measurement:
•  Plug the measuring adapter between the electrical consumer and power socket. Ensure the 
firm fit of the power plugs.
•  Observe the instruction manual of your current clamp or multimeter. 
•  Place the current clamp around the measuring chamber L or PE for the current to be 
determined. The current clamp must be completely closed to avoid measuring errors.
•  If you use a multimeter for voltage measurement proceed in the following order:
 - Connect the safety measuring cables to the digital-multimeter according to the instruction 
manual.
 - Use only suitable safety measuring cables.
 - Open the cover of the measuring adapter and plug the measuring cables into the sockets 
on which you want to measure a voltage.
•  Remove the current clamp from the measuring chamber and the plugs of the safety 
measuring cables from the measuring sockets after carrying out all measurements. Close 
the cover for the measuring sockets. 
•  Remove the measuring adapter from the power line.
The following measuring values can be determined:
Measuring chamber
Reading
L
Current via L respectively N
PE
Leakage current (stray current)