ACQUISITION - OFFSET, OFST
Description
The OFFSET command allows adjustment of the vertical offset of the specified input channel.
The OFFSET? query returns the DC offset value of the specified channel.
The OFFSET? query returns the DC offset value of the specified channel.
The maximum ranges depend on the fixed sensitivity setting. If an out-of-range value is
entered, the oscilloscope is set to the closest possible value.
entered, the oscilloscope is set to the closest possible value.
NOTE: The probe attenuation factor is not taken into account in offset adjustments.
Also, the unit V is optional.
Also, the unit V is optional.
Command Syntax
<channel>:OFfSeT <offset>
<channel>:= {C1, C2, C3, C4}
<offset>:=
Refer to datasheet specification for your Teledyne LeCroy oscilloscope at
www.teledynelecroy.com.
Query Syntax
<channel>:OFfSeT?
Response Format
<channel>:OFfSeT <offset>
Example
The following instruction sets the offset of Channel 2 to -3 V:
C2:OFST -3V