DS1000Z Quick Guide 
24 
Troubleshooting
 
 
The commonly encountered failures and their solutions are listed 
below. When you encounter those failures, please solve them 
following the corresponding steps. If the problem remains still, 
please contact 
RIGOL and provide your device information 
(acquisition method: Utility  System  System Info). 
 
1.  The screen is still dark (no display) after power on:   
(1)  Check if the power is correctly connected. 
(2)  Check whether the fuse is burned out. If the fuse needs 
to be changed, please use the specified fuse. 
(3)  Restart the instrument after finishing the above 
inspections. 
(4)  If it still does not work correctly, please contact 
RIGOL
 
2.  The signal is sampled but no waveform of the signal is 
displayed: 
(1)  Check if the probe is correctly connected to the 
oscilloscope and the item under tested.   
(2)  Check if there are signals generated from the item under 
test (you can connect the probe compensation signal to 
the problematic channel to determine which has 
problems, the channel or the item under test). 
(3)  Resample the signal. 
 
3.  The voltage amplitude measured is greater or lower 
than the actual value (note that this failure usually 
only occurs when probe is used):   
Check whether the attenuation coefficient of the channel