Ziehl MINIPAN 300 AC Digital panel-mounted measuring device MINIPAN 300 AC 1 mV - 500 V/AC / 1 mA - 1 A Assembly dimensi D 440320.CO データシート

製品コード
D 440320.CO
ページ / 12
date / name ) : 06.09.07 WL / Fz/Fu 
Z. Nr.: 12110-0710-01 
Page  11  of  12 
Type: Minipan 300 
ZIEHL industrie-elektronik GmbH + Co KG, Daimlerstr.13, D-74523 Schwäbisch Hall, Tel.: +49 791 504-0, Fax: -56, e-mail: info@ziehl.de 
Technical data
 
 
Power Supply 
  
Rated supply voltage Us  
AC/DC 24-240 V 
  Tolerance DC 
DC 20 - 297 V (0,85 x 24 V...1,35 x 220 V) 
  Tolerance AC 
AC 20 - 264 V (0,85 x 24 V...1,1 x 240 V) 
  Power consumption 
< 3 VA 
  Frequency 
48...62 Hz
 
 
Measuring Inputs
  
insulated from supply-voltage 
    
(connect 1 input only) 
  DC-Meter 
± 
300 mV / 120 kΩ / max. ±2,5 V 
  Measuring range / resistance of input  ± 
10.00 V / 1 MΩ / max. ±50 V 
  / overload capacity 
± 500.0 V / 3 MΩ / max. ±600 V 
   
± 100.0 V / 3 MΩ / max. ±600 V 
   
± 20.00 mA / Shunt 15 Ω/ max. ±100 
   
± 
1.00 A / Shunt 150 mΩ / max. ±2 A 
  AC-Meter 
150 mV / 900 Ω / max. 2,5 V 
  Measuring range / resistance of input 
10.00 V / 100 kΩ / max. 50 V 
  / overload capacity 
50.0 V / 1 MΩ / max. 60 V 
   
500.0 V / 3 MΩ / max. 600 V 
   
1.00 A / Shunt 150 mΩ / max. 2 A 
  Measuring time AC/DC 
< 400 ms + delay of display ddis 
  Pt 100- Meter 
- 199,9 ... + 850,0 °(= -328 ... +1563 °F) 
  Sensor-connection 
Pt 100, 2- or 3-wire 
  Wire-resistance 3-wire 
max. 3 x 50 Ω 
  Measuring time temperature 
< 400 ms + delay of display ddis  
 
Accuracy  
  Resolution 
+9999 / -1999 
  Error (of full measuring range) 
 
  DC voltage and current 
± 0,1 % ± 1 Digit 
  AC voltage and current 
± 0,5 % ± 1 Digit 
  Temperature factor 
± 0,02 % / Kelvin 
  Total error  
  At Pt 100- Meter 
± 0,3 % of value ± 0,5 K 
  Temperature factor 
± 0,03 °C / K 
 
Test conditions 
EN 50178 / EN 61010-1 
   
EN 61326-1 industrial applications 
  Measuring category 
CAT II  600 V DOUBLE INSULATION 
   
CAT III  300 V  DOUBLE INSULATION 
  Contamination level 
  On-time 
100 % 
  Rated ambient temperature range 
-20 °C ... +60 °C 
   
EN 60068-2-1 dry heat 
  Vibration resistance EN 60068-2-6 
2…25 Hz ±1,6 mm  
   
25 ... 150 Hz 5 g