To Set up the Vertical System
To Set the Channels
Each channel of DS1000B series digital oscilloscope has an independent operation
menu and it will pop-up after pressing any button among CH1, CH2, CH3, CH4. The
settings of all items in the menu are shown in the table below.
menu and it will pop-up after pressing any button among CH1, CH2, CH3, CH4. The
settings of all items in the menu are shown in the table below.
Figure 2-1 Table 2-1 The Channel menu (Page 1/2)
Menu
Settings
Comments
Coupling
AC
DC
GND
DC
GND
Restraint the DC component of the
input signal and passes the AC
component with frequency higher
than 5Hz.
Passes
input signal and passes the AC
component with frequency higher
than 5Hz.
Passes
both
AC
and
DC
components of the input signal
Disconnect the input signal.
Disconnect the input signal.
BW Limit
ON
OFF
OFF
Limit the channel‟s bandwidth to
20MHz to reduce display noise.
Get full bandwidth.
20MHz to reduce display noise.
Get full bandwidth.
Probe
0.001X
.
.
1000X
1000X
Set up the oscilloscope‟s probe
attenuation factor to make the
vertical scale readout correct.
Digital filter
Set up digital filter (See table 2-4).
1/2
Go to the next menu page (The
followings are the same, no more
explanation).