Lascar Electronics EL-USB-CO Carbon Monoxide Datalogger EL-USB-CO データシート

製品コード
EL-USB-CO
ページ / 3
Page 1 of 1
FEATURES
DATA LOGGER SET-UPS
• Logger Name
• Logging Rate (10s, 30s, 1m, 5m)
• Warning Threshold
• Start Date and Start Time
ORDERING INFORMATION
Stock Number
Standard Data Logger
EL-USB-CO
Replacement Battery
BAT 3V6
(Data Logger, Software on CD,
Battery)
www.lascarelectronics.com
This data logger measures and stores up to 32,510 Carbon Monoxide (CO) readings over a 0 to 1000 ppm measurement range and -
10 to +40°C (14 to +104°F) operating temperature range. The user can easily set up the logging rate and start time, and download
the stored data by plugging the module straight into a PC's USB port and running the purpose designed software under Windows 98,
2000 or XP. The data can then be graphed, printed and exported to other applications. The data logger is supplied complete with a
long-life lithium battery. Data is stored in non-volatile memory and is not lost when the battery is empty. A bright red LED will flash
and a buzzer can sound when a preset warning level has been exceeded.
• 0 to 1000 ppm CO measurement range
• Stores up to 32,510 Measurements
• User-Programmable Warning Threshold
with Bright Visual and Audible Warning
• USB Interface for Set-up and Data Download
• Expected sensor life: 4 years
• Replaceable Internal Lithium Battery
WINDOWS CONTROL SOFTWARE
Easy to install and use, the control software runs under
Windows 98, 2000 and XP (Home and Professional
Editions). It allows the user to set up and download any
EL-USB-CO. The latest version of the control software may
be downloaded from
.
www.lascarelectronics.com
EL-USB-CO
Carbon Monoxide (CO) Data Logger with USB Interface
IMPORTANT SAFETY WARNING:
DO NOT ATTEMPT TO USE THE EL-USB-CO WITHOUT READING AND UNDERSTANDING THE DOCUMENTATION.
CARBON MONOXIDE MONITORING IS A COMPLEX PROCEDURE. IF THERE IS ANY DOUBT AS TO THE SAFETY OR
APPLICATION - STOP! CONSULT A QUALIFIED EXPERT.
The EL-USB-CO is a professional measuring instrument, designed to record Carbon Monoxide (CO) gas levels for later
analysis. It is not an alarm/safety/warning device and should not be used in this capacity for the following reasons:
1.
2.
3.
4.
The lights and sound emitted by the EL-USB-CO to indicate a high level of CO do not provide adequate warning to those in the
surrounding area that a high level of CO has been recorded.
The battery life of the EL-USB-CO during audible and visual indication of a high level of CO will rapidly diminish and will therefore only
be active for a short period of time.
The EL-USB-CO takes a sample reading at intervals as set by the user for recording/monitoring/analysis purposes. Dangerous levels of
CO will not register with the device until a sample reading is taken.
Due to the limitations of the device as a professional instrument, it is left to the user to set the light and sound alarms indicating the
presence of company Therefore the instrument can be set up to provide NO form of warning.
Carbon Monoxide is poisonous. It is clear, colorless gas that has no smell or taste and can be lethal even in very low concentrations. It can
injure or kill people and animals who are not adequately warned of its presence.
In areas where the safety of humans and animals is a concern we insist that a domestic or industrial CO detector/alarm, designed to a
nationally approved quality standard, with features such as a high decibel alarm or siren and a clear visual warning, specifically designed
to provide adequate warning of dangerous CO levels, be installed, used and maintained.
Specifications liable to change without prior warning
EL-USB-CO
Issue 2
April/2006
M.C.
Applies to EL-USB-CO/2