Gossen Metrawatt Metrahit Iso Insulation measuring device, CAT III 600 V Metrahit Iso ユーザーズマニュアル
製品コード
Metrahit Iso
GMC-I Gossen-Metrawatt GmbH
61
Technical Data
Electrical Safety
Safety class
II per EN 61010-1:2001/VDE 0411-1:2002
Measuring category
CAT II
CAT III
Nominal voltage
1000 V
600 V
Fouling factor
2
Test voltage
5.2 kV~ per EN 61010-1:2001/VDE 0411-
1:2002
1:2002
Fuses
Fuse link
FF 10 A/1000 V AC/DC;
10 mm x 38 mm; switching capacity:
30 kA at 1000 V AC/DC;
protects the current measurement input in
the 300
10 mm x 38 mm; switching capacity:
30 kA at 1000 V AC/DC;
protects the current measurement input in
the 300
μA through 10 A ranges
Electromagnetic Compatibility (EMC)
Interference emission EN 61326: Oct. 2006, class B
Interference immunity EN 61326: Oct. 2006, appendix A
Interference immunity EN 61326: Oct. 2006, appendix A
IEC 61000-4-2: Dec. 2001
Feature B
Feature B
8 kV atmospheric discharge
4 kV contact discharge
4 kV contact discharge
IEC 61000-4-3: Dec. 2006
Feature B:
Feature B:
3 V/m
Data Interface
Type
Optical via infrared light through the housing
Data transmission
Serial, bidirectional (not IrDa compatible)
Protocol
Device specific
Baud Rate
38,400 baud
Functions
– Select/query measuring functions
and parameters
– Query momentary measurement data
The USB
⏐
X-
TRA
plug-in interface adapter (see accessories) is
used for adaptation to the PC’s USB port.
Internal Measured Value Storage
Memory capacity
4 MBit / 540 kB for approx. 15,000
measured values with indication of date
and time
measured values with indication of date
and time
Mechanical Design
Housing
Impact resistant plastic (ABS)
Dimensions
200 x 87 x 45 mm
(without protective rubber cover)
(without protective rubber cover)
Weight
Approx. 0.35 kg with batteries
Protection
Housing: IP 54
(pressure equalization by means of the
housing)
(pressure equalization by means of the
housing)
Table Excerpt Regarding Significance of the IP Code
IP XY
(1
(1
st
digit X)
Protection against
foreign object entry
foreign object entry
IP XY
(2
(2
nd
digit Y)
Protection against the
penetration of water
penetration of water
4
≥ 1.0 mm dia.
4
splashing water
5
dust protected
5
water jets