APPENDIX C/ Battery Specifications and Replacement Procedure
Battery Pack Safety Precautions
WARNING
DO NOT intentionally short circuit the battery pack. The batteries are capable of 
providing hazardous output currents if short circuited. The Energy Platform is 
equipped with an internal battery charger circuit. Do not attempt to charge the 
batteries with an external charger other than the Dranetz-BMI battery charger, 
since improper charging could cause battery explosion.
ADVERTENCIA
NO ponga intencionalmente la baterÌa en cortocircuito. Las baterÌas son capaces 
de proporcionar corrientes de salida peligrosas si est·n en cortocircuito. La Energy 
Platform est· equipada con un circuito interno cargador de baterÌas. No intente 
cargar las baterÌas con un cargador externo que no sea el cargador de baterÌas 
Dranetz-BMI, puesto que la carga indebida podrÌa hacer que explote la baterÌa.
AVERTISSEMENT
NE PAS court-circuiter délibérément le bloc-batterie. Lors díun court-circuit, les 
batteries risquent díémettre des courants effectifs dangereux. Energy Platform 
posséde un circuit de chargeur de batterie intégré. Ne pas tenter de charger les 
batteries au moyen díun chargeur externe autre que le chargeur de batterie 
Dranetz-BMI, car un rechargement fautif pourrait entraÓner líexplosion de la 
batterie.
WARNUNG
Die Batterien dürfen NICHT kurzgeschlossen werden. Im Falle eines 
Kurzschlusses kˆnnen die Batterien lebensgefährliche Ausgangsstrˆme leiten. 
Energy Platform ist mit einem internen Batterieladegerät ausgestattet. Die 
Batterien sollten nur mit dem Ladegerät von Dranetz-BMI geladen werden. Die 
Verwendung eines externen Ladegeräts kann zu einer Explosion der Batterien 
führen.  
Battery safety 
precautions
Observe the following precautions when inspecting or replacing the battery pack:
• Do not attempt to replace individual batteries of the pack or substitute other battery 
types.
• Do not dispose of battery in fire.
• Dispose of a used battery promptly in accordance with local Environmental 
Protection Agency (EPA) regulations.
• Visually inspect the battery pack for corrosion.
The batteries have a safety pressure vent to prevent excessive gas build-up and 
corrosion indicates that venting has occurred. Possible causes of venting are: a 
defective charger, excessive temperature, excessive discharge rate, or a defective 
cell.
If corrosion is excessive, the battery pack may require replacement (contact Dranetz-