15 Specifications 
Display range 
maximal -19999...19999 digit, depends on used sensor 
Number of channels 
Suitable probes 
GMSD 
… - K51 and MSD … sensors with connection cable MSD-K51 
Available ranges/resolutions  
ranging from  -
1.999 … 2.500 mbar / 0.001 mbar  to  0.0 … 1000 bar / 1 bar 
Connections 
sensor 
7-pole bayonet connector 
output /  
ext. supply 
4-pole connector for serial interface and supply (USB Adapter USB 5100) 
Display units 
depends on meas. range selection (depends on connected sensor): 
mbar, bar, Pa, kPa, MPa, mmHg, inHg, PSI, mH
2
User unit 
freely adjustable 
Measuring frequency 
4 / s  or 1000 / s 
Accuracy 
± 0.1 % FS ± 1 digit 
Working conditions 
-25 ... 50 °C; 0 ... 95 %RH (non condensing) 
Storage conditions 
-25 ... 70 °C 
Display 
4 ½ digit 7-segment display, illuminated (white) 
Correction function 
offset / slope (via menu),  
if offset/slope value not equal zero: "corr" is displayed 
Additional functions 
min- / max- / hold- function 
Averaging 
 
adjustable
, 1 … 120 seconds 
Housing 
protection class  IP65 and IP67 
 
dimensions 
L*W*H [mm] 
160 * 86 * 37 incl. protective covering, 
approx. 250 g incl. batteries and protective covering 
Power supply 
 
2*AAA-batteries (included in delivery)  
current 
consumption 
2.0 mA (at Out = Off, equivalent to 500 h),  
illumination ~10mA (turns off automatically) 
Battery-change indicator 
automatically if battery exhausted "bAt", warning "bAt" flashing 
Auto-off-function 
Device will be automatically switched off if no key is pressed/no interface 
communication takes place for the time of the power-off delay.  
The power-off delay can be set to values between 1 and 120 min.; it can be 
completely deactivated. 
EMC 
The device corresponds to the essential protection ratings established in the 
Regulations of the Council for the Approximation of Legislation for the member 
countries regarding electromagnetic compatibility (2004/108/EG) 
Additional fault: <1%