4 Device Configuration
Some menu items will be shown depending on the actual device configuration (e.g. there are
some items disabled when the logger contains data). Please note the hints by the menu items.
For configuration of the device press "Menu"-key (key 4) for 2 seconds, the main menu will be shown (main display:
“SEt”). Choose the desired menu branch by pressing the "Menu"-key (key 4). By pressing “
“SEt”). Choose the desired menu branch by pressing the "Menu"-key (key 4). By pressing “
” (key 3) the referring
parameters can be chosen. The referring values are changed by pressing the keys "
" (key 2) or "" (key 5) (Choice
of parameter: “
”). Pressing “Menu” (key 4) again will jump back to the main menu selection and stores the settings.
Use key "Store/
" (key 6) to leave configuration..
Menu
Parameter Values
Meaning
key Menu
key
key
or
p.r.t.
Set Sort
Set Sort: limitation of the material selection
off:
Unrestricted material selection via key 2 and 5
*
1...8:
Material selection in-between 1 up to 8 selectable materials
....
selectable materials (not available if Sort = off). Select the desired
material that should be available during the measure via key 2 and 5.
material that should be available during the measure via key 2 and 5.
*
Set
Conf
Conf
Set Configuration: Generic Settings
Arrow bottom left
points to
points to
“%u”
Moisture display = moisture content [%u]
*
Arrow bottom left
points to
points to
“%w”
Moisture display = wet-basis moisture content [%u]
°C
All temperature values are in degrees Celsius
°F
All temperature values are in degrees Fahrenheit
oFF
Atc off: temperature input for compensation via keys
*
*
*
5.4
on
Atc on: temperature compensation via internally measured temperature
or external probe
or external probe
oFF
Auto HLD off: continuous measuring.
*
5.3
on
Auto-HLD on: when reaching a stable measuring result, this will be
frozen with-HLD. When pressing the store-key a new measure will be
initiated. If logger is switched on (‚Func CYCL‘, ‚Func Stor‘): device
works like setting would be “auto-HLD off”
frozen with-HLD. When pressing the store-key a new measure will be
initiated. If logger is switched on (‚Func CYCL‘, ‚Func Stor‘): device
works like setting would be “auto-HLD off”
oFF:
Average value calculation off
5.6
on:
Average value calculation on: Average value out of 3 sequential
measuring is calculated and displayed
measuring is calculated and displayed
1...120
Power-off delay in minutes.
Device will be automatically switched off as soon as this time has
elapsed if no key is pressed/no interface communication takes place
Device will be automatically switched off as soon as this time has
elapsed if no key is pressed/no interface communication takes place
oFF
Power-off function inactive (continuous operation, e.g. mains operation)
oFF
Function of the output: No output function, lowest power consumption
SEr
Output is serial interface
dAC
Output is analogue output 0...1V
01, 11 .. 91
Base Address when Output = Serial Interface :
Base address of device for interface communication.
Base address of device for interface communication.
0.0 ... 100.0%
Enter desired moisture value at which the analogue output potential
should be 0V
should be 0V
0.0 ... 100.0%
Enter desired moisture value at which the analogue output potential
should be 1V
should be 1V
Set Logg
Set Logger: Configuration Of Logger Function
CYCL
Cyclic
: logger function ‚cyclic logger‘
*
Stor
Store
: logger function ‚individual value logger‘
oFF
no logger function
0:30 ... 60:00
Cycle time of cyclic logger [minutes:seconds]
*
Set CLOC
Set Clock: Setting Of Real Time Clock
HH:MM
Clock: Setting of time hours:minutes
YYYY
Year
DD.MM
Date: day.month
The settings will be set to the settings ex works, if keys ‘Set‘ and ‘Store‘ are pressed
simultaneously for more than 2 seconds.
simultaneously for more than 2 seconds.
(*)
If the logger memory contains data already, the menus/parameters marked can not be called!
If these should be altered the logger memory has to be cleared before!
If these should be altered the logger memory has to be cleared before!
If the manual logger contains data (Logger: ‘Func Stor’), the first menu displayed will be: ‘rEAd Logg’
please refer to chapter 7.1
please refer to chapter 7.1