4  Device Configuration 
 Some menu items will be shown depending on the actual device configuration (e.g. there are 
some items disabled when the logger contains data). Please note the hints by the menu items. 
For configuration of the device press "Menu"-key (key 4) for 2 seconds, the main menu will be shown (main display: 
“SEt”). Choose the desired menu branch by pressing the "Menu"-key (key 4). By pressing “
” (key 3) the referring 
parameters can be chosen. The referring values are changed by pressing the keys  "
" (key 2) or "" (key 5) (Choice 
of parameter: “
”). Pressing “Menu” (key 4) again will jump back to the main menu selection and stores the settings. 
Use  key "Store/
" (key 6) to leave configuration..  
Menu 
Parameter  Values 
Meaning 
 
 
key Menu 
key

key
 
or
 
 
 
 
p.r.t. 
Set Sort 
 
Set Sort: limitation of the material selection
 
 
 
 
off
Unrestricted material selection via key 2 and 5 
*
 
 
1...8: 
Material selection in-between 1 up to 8 selectable materials 
....
 
selectable materials (not available if Sort = off). Select the desired 
material that should be available during the measure via key 2 and 5. 
*
 
 
Set 
Conf 
 
 
Set Configuration: Generic Settings
 
 
 
 
Arrow bottom left 
points to
 
“%u” 
Moisture display = moisture content [%u] 
 
Arrow bottom left 
points to
 
“%w” 
Moisture display = wet-basis moisture content [%u] 
 
°C 
All temperature values are in degrees Celsius 
 
 
°F 
All temperature values are in degrees Fahrenheit 
 
oFF 
Atc off: temperature input for compensation via keys 

5.4 
on 
Atc on: temperature compensation via internally measured temperature 
or external probe 
 
oFF 
Auto HLD off: continuous measuring. 
* 
5.3 
on 
Auto-HLD on: when reaching a stable measuring result, this will be 
frozen with-HLD. When pressing the store-key a new measure will be 
initiated. If logger is switched on (‚Func CYCL‘, ‚Func Stor‘): device 
works like setting would be “auto-HLD off”
 
 
oFF: 
Average value calculation off 
 
5.6 
on: 
Average value calculation on: Average value  out of 3 sequential 
measuring is calculated and displayed  
 
1...120 
Power-off delay in minutes.  
Device will be automatically switched off as soon as this time has 
elapsed if no key is pressed/no interface communication takes place 
 
 
oFF 
Power-off function inactive (continuous operation, e.g. mains operation) 
 
oFF 
Function of the output: No output function, lowest power consumption 
 
SEr 
Output is serial interface 
dAC 
Output is analogue output 0...1V 
 
01, 11 .. 91 
Base Address when Output = Serial Interface : 
 Base address of device for interface communication. 
 
 
0.0 ... 100.0% 
Enter desired moisture value at which the analogue output potential 
should be 0V 
 
 
0.0 ... 100.0% 
Enter desired moisture value at which the analogue output potential 
should be 1V 
 
Set Logg 
 
Set Logger: Configuration Of Logger Function 
 
 
 
CYCL 
Cyclic
: logger function ‚cyclic logger‘ 
Stor 
Store
: logger function ‚individual value logger‘ 
oFF 
no logger function 
 
0:30 ...  60:00 
Cycle time of cyclic logger [minutes:seconds] 
Set CLOC 
 
Set Clock: Setting Of Real Time Clock 
 
 
 
HH:MM 
Clock: Setting of time  hours:minutes 
 
 
 
YYYY 
Year 
 
DD.MM 
Date:                            day.month 
The settings will be set to the settings ex works, if keys ‘Set‘ and ‘Store‘ are pressed  
 simultaneously for more than 2 seconds. 
(*) 
If the logger memory contains data already, the menus/parameters marked can not be called! 
If these should be altered the logger memory has to be cleared before! 
If the manual logger contains data (Logger: ‘Func Stor’), the first menu displayed will be: ‘rEAd Logg’  
please refer to chapter 7.1