 Operating Manual GMH 3851 
Page 9 of 22 
 
 
_____________________________________________________ 
_____________________________________________________________________________ 
 
5.6.3  Measuring 
Split the log:  
Use an suitable axe for splitting. Fast working splitting machines can produce heat which may falsify the 
measuring. 
 
Preparation of core temperature measuring of log:  
Drive in the impact electrode, put temperature probe into the generated hole. 
 
Three measuring values have to be taken at the freshly split surface, 
 
 
 
2 measuring spots should have 5 cm distance 
from cut side, the third should be placed in the 
center.  
Drive in the needles sufficiently deep( > 5mm) and 
across the fibre direction 

 
 
 
 
 
Measuring procedure: 
 
Step 
Action 
Description 
Display 
Measuring 1 
Drive in needles 
 
 
Start measuring 
press
 
Measuring value 1 will be 
automatically taken 
 
Measuring stable:  
 
 
Instrument is ready for 
next step 
 
Measuring 2 
Drive in needles 
 
 
Start measuring 
press
 
Measuring value 2 will be 
automatically taken 
 
Measuring stable:  
 
 
Instrument is ready for 
next step 
 
Measuring 3 
Drive in needles 
 
 
Start measuring 
press
 
Measuring value 3 will be 
automatically taken 
 
Measuring stable:  
 
 
Average out of the 3 will 
be displayed 
e.g. 
 
Return to continous 
measuring display 
press