Extech Sound level-measuring apparatus, Noise-measuring apparatus SDL600 データシート

製品コード
SDL600
ページ / 3
 
Extech  SW  Protocol   (All Versions) 
 
 
 
 Meter   -  RS232 Protocol 
 
RS232 Format:  Baud Rate: 9600, Parity: None,  Data Bits: 8,  Stop Bits: 1,  Flow       
                          Control: XOn/XOff  (NONE) 
 
16 digit data stream format: 
 
D15 D14 D13 D12 D11 D10
D9 D8 D7 D6 D5 D4 D3 D2 D1 D0 
 
 
D0 End 
Word 
D1 to D4 
Upper Display Reading; D1=LSD, D4=MSD 
D5 to D8 
Lower Display Reading; D5=LSD, D8=MSD 
D9 
Decimal Point (DP) for Upper Display 
0= No DP, 1= 1 DP, 2 = 2 DP, 3 = 3 DP 
D10 
Decimal Point (DP) for Lower Display 
0= No DP, 1= 1 DP, 2 = 2 DP, 3 = 3 DP 
D11 & D12 
Annunciator for Upper Display 
00 = No Symbol 
07 = mg/L 
14 = mS 
01= °C 
08 = m/s 
15 = Lux 
02 = °F 
09 = Knots 
16 = FT-cd 
03 = % 
10 = Km/hr 
17 = dB 
04 = %RH 
11 = Ft/min 
18 = mV 
05 = %PH 
12 = mile/hr 
 
06 = %O2 
13 = uS 
 
D13 
Anunciator for Lower Display      
0 = No Symbol     1 = C     2 = F 
D14 
Reading Polarity for the Display 
0 = Both Upper & Lower Display Value are “+”. 
1 = Upper “-“, Lower “+” 
2 = Upper “+”, Lower “-“ 
3 = Both Upper & Lower Display Values Are “-“ 
D15 Start 
Word 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Meter 
End 
     PC 
End 
3.5mm phone plug 
 
 
 
       9 pin D connector 
RS232 Cable 
Tip  to Pin 2 
Sleeve (ground)  to Pin 5