 Operating manual GTH 200 air 
page 5 of 8 
 
 
_____________________________________________________ 
_____________________________________________________________________________ 
 
3.5  Display of min-/max- value memory 
For all measuring values the lowest and highest measured value since the switch-on of the device will be 
stored. 
Display MIN value (Lo): 
press 
 shortly 
display switching between 
„Lo‟ and MIN value  
Display MAX value (Hi): 
press 
 again 
display switching between 
„Hi‟ and MAX value 
Return to current value: 
press 
 again  
current measured value is displayed  
Delete MIN/MAX value: 
press 
 for 2s 
MIN and MAX values are deleted.  
  
„CLr‟ (Clear) is displayed shortly 
After switching-off and turning-on again all min-/max- values will be deleted. 
3.6  Hold 
– function 
When the key 
  key is pressed, the current measured values will be 
„frozen‟ (display symbol: HLD). As 
soon as the key 
 is pressed again, the device measures again the normal way.  
3.7  System messages 
   ER.   1 
measuring range has been exceeded, measuring value too high 
   ER.  2 
meas. value has fallen below permissible range, measuring value too low 
   ER.  7 
system error - the device has detected a system fault  
(defective or far outside allowable ambient temperature range) 
 
 
If 
“BAT” is displayed at the left side of display, the battery is weak, measuring 
can be continued for a short period. 
   BAT  
The battery is used up and needs to be replaced. Measuring is no more possible. 
4  Configuration of the device 
To configure the device functions proceed as follows:  
  Switch off the device. 
  Press 
 and keep button pressed. Switch on device (press 
 shortly).  
Release 
 
not before the first parameter “
P_OF
” is displayed. 
 
Adjust current parameter with “up” 
 
or “down” 
  Switch to the next parameter with button 
Parameter 
  Value 
  Description 
 
Auto power-off delay  
ex works: 20 min.
 
 120 
Auto power-off delay in minutes. Device will be automatically switched off as 
soon as this time has elapsed if no key is pressed. (possible values: 1..120 min) 
OFF 
The auto power off function is deactivated (continuous operation). 
 
Temperature unit   
ex works: °C
 
° C / °F 
Temperature in °C / in °F 
 
Restore factory defaults 
NO YES 
Abort / restore factory defaults 
Pressing 
 after adjusting the last parameter will save the settings and restart the device 
(segment test). 
Please note: 
If no key is pressed for more than 2 minutes the configuration will be aborted. All changes 
will be discarded!