Microchip Technology ARD00330 データシート

ページ / 62
Meter Calibration
© 2011 Microchip Technology Inc.
DS51964A-page 33
2.4.3
RMS Current, RMS Voltage, Signal Flow and Calibration
The RMS current and voltage outputs require a two-point calibration at configurations
C1 and C5. The automated software performs these calibrations, suggested on the
calibration values entered in the text boxes, as shown in Figure 2-16.
Table 2-2 and Figure 2-18 represent the registers that are set for RMS current and 
voltage calibration.
TABLE 2-2:
RMS CURRENT, RMS VOLTAGE, CALIBRATION REGISTERS
FIGURE 2-18:
RMS Current and Voltage Flow.
Register
Equation
Configurations Required
OFFSET_V_RMS
C1, C5
OFFSET_I_RMS
C1, C5
GAIN_V_RMS
C1 ONLY
GAIN_I_RMS
C1 ONLY
X
2
Σ
OFFSET_I_RMS:16
X
2
Σ
OFFSET_V_RMS:16
ADC
ADC
CURRENT
VOLTAGE
RMS Current
RMS Voltage 
X
V
X
A
GAIN_V_RMS:16
GAIN_I_RMS:16
I_RMS:16
RAW_I_RMS:16
V_RMS:16
RAW_V_RMS:16