The zoom factor is displayed above the grid: 
 
The zoom magnification factor is the ratio of the timebase of the zoom trace to that of 
the input waveform. 
Zoom delay represents the portion of the input waveform being zoomed. As you turn 
the Horizontal delay knob, this value becomes positive or negative depending on 
whether the zoom is left (+) or right (-) of center. 
The symbols M and Z indicate that the Main grid is on top and the Zoom grid is on the 
bottom of the scope display. 
Replay Mode 
Replay mode provides a way to scroll backwards in time to view past acquisitions. 
 
The number of sweeps that can be stored depends on the Max Memory Length 
selected in the “Horizontal” menu. Replay Mode does not function under the following 
conditions: 
  in Roll Mode 
  in Average Mode 
  in Equivalent Sampling (RIS) mode 
 
To access Replay mode and scroll 
through waveforms, press 
S
TOP
/S
INGLE
 to stop acquisitions, 
then press the I
NTENSITY
/R
EPLAY
 
knob. The Replay LED lights to 
confirm Replay Mode.