8 Configuration of the instrument
To configure the instruments according to Your needs proceed like follows::
Switch off instrument.
Switch on and press mode-key during the segment test ( ), until the display shows
the first parameter ‚P_oF„
If a parameter should be edited, press key up or down ( ), the setting of the parameter
will be shown and can be changed via up/down keys and receipt the changed parameter with
Jump to the next parameter with key
parameter
value
meaning
key
keys
Auto Power-Off
factory setting: Auto
Auto Power-Off in minutes.
If there won‟t be pressed, the device will be
switched off automatically to save battery power (adjustable range 1 to 120
min)
min)
automatic power-off is deactivated (continuous operating)
Unit of display
factory setting: cond µS/cm
cond µS/cm
Measuring of conductivity in unit µS/cm
rESi MOhm*cm Measuring of resistivity in unit MOhm*cm
Range of display
factory setting: Auto
Auto
Measuring with auto-range
100.0 µS/cm
If Unit = cond µS/cm: conductivity range 0.0 ... 100.0 µS/cm
20.00 µS/cm
“ : conductivity range 0.00 ... 20.00 µS/cm
2.000 µS/cm
“ : conductivity range 0.000 ... 2.000 µS/cm (no auto-range)
.2000 MOhm*cm If Unit = rESi MOhm*cm: resistivity range 0.0100 ... 0.2000 MOhm*cm
2.000 MOhm*cm
2.000 MOhm*cm
“ : resistivity range 0.010 ... 2.000 MOhm*cm
20.00 MOhm*cm
“ : resistivity range 0.01 ... 20.00 MOhm*cm
Unit of temperature display
factory setting: °C
°C
temperature is displayed in °C
°F
temperature is displayed in °F
temperature compensation (only for µS/cm and mS/cm)
factory setting: nLF
OFF
No temperature compensation of conductivity measuring
nLF
non-linear temperature compensation for natural water acc. to EN 27888
(DIN 38404) to measure ground water, surface water, drinking water and
purest water
(DIN 38404) to measure ground water, surface water, drinking water and
purest water
Lin
Linear temperature compensation with adjustable coefficient (t.Lin)
nA.CL
Temperature compensation of weak NaCl-solutions acc. to DIN EN 60746-3
Reference temperature of temp. compensation (only for t.Cor <> oFF)
factory setting: 25 °C
25 °C / 77 °F
Reference temperature 25 °C / 77 °F
20 °C / 68 °F
Reference temperature 20 °C / 68 °F
Compensation coefficient (only for t.Cor = Lin) factory setting: 1.000
0.300 .. 5.000
Temperature compensations coefficient in % / K.
The factor has to be evaluated by experiment or resolved from literature.
The factor has to be evaluated by experiment or resolved from literature.
Restore to factory settings
Settings are kept
Restoring device to factory settings
Pressing again stores the settings, the instruments restarts (segment test)
Please note: If there is no key pressed within the menu mode within 2 minutes, the configuration
will be cancelled, the entered settings are lost!