modes, mode parameters, and durations.  They are useful for executing a set of tests on a device, 
then displaying whether the tests passed or failed.  We will illustrate how to use test files by a short 
example.
Suppose we have a small AC to DC power supply (a "wall-wart") and we want to set up an 
acceptance test for a number of these devices.  Our test will consist of two steps:
1. Set the DC load to constant current mode to draw the rated current of 0.35 A from the 
device.  The output voltage of the device at the rated current must be between 4.4 volts and 
4.6 volts.
2. When the device operates into a short, the supplied current must be larger than 2.0 A.
Keys
Display
Shift + Menu
:CONFIG
 
▽ ▽
:LIST SET
Enter
:MODE SET
     
▽ ▽ ▽ ▽
:EDIT TEST FILE
Enter
MAX CURR= 3.000A
  Set the maximum current to 2.5 A
2.5 Enter
MAX VOLT= 18.00V
  Set the maximum voltage to 5 V
5 Enter
MAX POWER= 150.00W
  Set the maximum power to 15 W
15 Enter
TEST COUNT= 6
  Our test will consist of two steps, so we enter 2.  You 
can enter up to 20 steps.
2 Enter
CONST CURRENT 
 Since our first test is in constant current mode, we'll 
just press Enter to select this mode.
Enter
SET 1= 0.210A
  This is the prompt for the constant current value.  We set 
it to 0.35 A.
.35 Enter
SHORT OFF 
 We can turn a short on or off at this step.  Here, we want 
short off, so we just press Enter.
Enter
READBACK V 
 We're being prompted for what parameter to read back 
and check.  We want voltage, so we press Enter.
Enter
MIN 1= 5.80V
  We want the minimum voltage read back to be at least 4.4 
V.
4.4 Enter
MAX 1= 6.15V
  The maximum value we'll allow is 4.6 V.
4.6 Enter
DELAY 1= 1.0<S>
  This delay time is how long to wait before making the 
readback measurement.  If you set it to 25.5 seconds, the test will halt at 
this point, requiring you to press Shift + Trigger to continue the test.  We'll 
use 1 second.
1 Enter
CONST CURRENT
  We're being prompted for the mode to use for the 
next step.  We'll again use constant current, so we'll press Enter (this step 
will also be a short, so the mode doesn't really matter).
Enter
SET 2= 5.000A 
 We'll set the current to 2.5 A, which is the maximum 
current we've allowed for this test..
8500 DC Load Series                    
Version:  September 2, 2009 