Ea Elektro Automatik EA Elektro-Automatik EA-ELR 9250-140 3U Electronic Load 0 - 140 A 0 - 250 Vdc 0 - 7000 W 33200406 データシート

製品コード
33200406
ページ / 160
Page 25
EA Elektro-Automatik GmbH
Helmholtzstr. 31-33 • 41747 Viersen
Germany
Fon: +49 2162 / 3785-0
Fax: +49 2162 / 16230
ELR 9000 Series
1.9.7  USB-Port Type B (Back side)
The USB-B port on the back side of the device is provided for communication with the device 
and for firmware updates. The included USB cable can be used to connect the device to a PC 
(USB 2.0 or 3.0). The driver is delivered on the included CD and installs a virtual COM port. 
Details for remote control can be found on the web site of Elektro-Automatik or on the included 
CD. A general programming introduction for the USB port is available (date: 01-15-2015) 
The device can be addressed via this port either using the international standard ModBus pro-
tocol or by SCPI language. The device recognises the message protocol used automatically.
If remote control is in operation the USB port has no priority over either the interface module 
(see below) or the analog interface and can, therefore, only be used alternatively to these. 
However, monitoring is always available.
1.9.8  Interface module slot
This slot on the back side of the device (only with standard models, units with option 
3W installed are different) is available for various modules of the AnyBus CompactCom 
(abbr.: ABCC) type of the IF-AB interface series. The following options are available 
(date: 01-15-2015): 
Article number Name
Description
35400100
IF-AB-CANO
CANopen, 1x Sub-D 9pole male
35400101
IF-AB-RS232
RS 232, 1x Sub-D 9pole male (null modem)
35400103
IF-AB-PBUS
Profibus DP-V1 Slave, 1x Sub-D 9pole female
35400104
IF-AB-ETH1P
Ethernet, 1x RJ45
35400105
IF-AB-PNET1P
ProfiNET IO, 1x RJ45
35400106
IF-AB-DNET
Devicenet, 1x Wago plug 5-pole
35400107
IF-AB-MBUS1P
ModBus TCP, 1x RJ45
35400108
IF-AB-ETH2P
Ethernet, 2x RJ45
35400109
IF-AB-MBUS2P
ModBus TCP, 2x RJ45
35400110
IF-AB-PNET2P
ProfiNET IO, 2x RJ45
The modules are installed by the user and can be retrofitted without problem. A firmware update of the device may 
be necessary in order to recognize and support certain modules.
If remote control is in operation the interface module has no priority over either the USB port or the analog interface 
and can, therefore, only be used alternately to these. However, monitoring is always available.
Switch off device before adding or removing modules! 
1.9.9  Analog interface
This 15 pole Sub-D socket on the back side of the device is provided for remote control of 
the device via analog signals or switching conditions. 
If remote control is in operation this analog interface can only be used alternately to the 
digital interface. However, monitoring is always available. 
The input voltage range of the set values and the output voltage range of the monitor val-
ues, as well as reference voltage level can be switched in the settings menu of the device 
between 0-5 V and 0-10 V, in each case for 0-100%.