2.3.6  Grounding of the DC input
Individually operated devices can always be grounded from the DC minus pole, i.e. can be directly connected to 
PE. The DC plus pole, however, if it is to be grounded, may only be so for input voltages up to 400 V, because the 
potential of the minus pole is shifted into negative direction by the value of the input voltage. Also see technical 
specification sheets in 1.8.2, item “Insulation”.
For this reason, for all models which can support an input voltage higher than 400 V grounding of the DC plus 
pole is not allowed. 
Do not ground the DC plus pole on any model with >400 V nominal voltage
If grounding one of the input poles ensure that no output pole of the source (e.g. power supply) 
is grounded. This could lead to a short-circuit! 
2.3.7  Connection of remote sensing
Both pins “NC” on the “Sense” terminal must not be connected!
The cross section of the sensing cables is noncritical. Recommendation for cables up to 5 m: 
use at least 0.5 mm²
Sensing cables should be twisted and laid close to the DC cables to damp oscillation. If neces-
sary, an additional capacitor should be installed at the source to eliminate oscillation
Sensing cables must be connected + to + and - to - at the source, otherwise the sense input 
In master-slave operation, the remote sensing should be connected to the master unit only
Figure 8 - Example for remote sensing wiring