Scala Fever thermometer SC29 flex 01244 データシート

製品コード
01244
ページ / 1
Classic waterproof clinical thermometer  
SC 29 flex with flexible measuring point 
 
Special features 
The  SC  29  flex  is  a  waterproof,  digital  fever  thermometer  with  a  flexible  measuring  tip.  It 
displays the highest temperature within 10 seconds. It has a memory function to store the last 
measured value. A fever alarm beeps when the temperature measured exceeds 37.5°.
  
 
Safety information 
 The SC 29 flex is suitable only for measuring human body temperatures. 
The  thermometer  contains  small  parts  (batteries,  etc)  which  could  be  swallowed  by  small 
children. Please keep the thermometer and its batteries out of the reach of small children. If the 
battery or other parts are swallowed you should seek medical help immediately. 
The  enclosed  batteries  may  not  be  recharged  or  reactivated  by  other  methods,  may  not  be 
dismantled, thrown into fires or short-circuited.  
Protect the SC 29 flex from direct sunshine and temperatures above 50°C.  
Never  drop  the  SC  29  flex,  do  not  bend  or  open  it  (with  the  exception  of  the  battery 
compartment for battery replacement). 
 
Microbial contamination 
Because a hundred percent disinfection is not possible under conditions at home, it is urgently 
recommended that every patient should use their own fever thermometer.  
 
Operating manual 
Press the I/O button to switch on. A short signal tone shows that the device is in the "switched 
on" mode. A segment test takes place simultaneously (checks all display segments).  
After  this,  if  the  temperature  measured  by  the 
temperature  sensor  is  less  than  32°C,  the  display 
shows  an  "L“    and  "°C“.  After  around  one  second,  the 
"°C“    starts  to  blink.  The  „L“  stands  for  low  (less  than 
32°C).  The  current  temperature  is  continuously 
displayed  during  measurement,  and  the  "°C"  blinks 
during the entire measurement time. The measurement 
is  completed  after  around  10  seconds  and  "°C"  stops 
blinking.  After  the  measurement  is  completed,  the 
temperature value measured remains in the display  
until the SC 29 flex is switched off. To turn the device off after use, press the I/O button briefly. 
The signal tone will again be sounded. If the device is not switched off, a signal tone is sounded 
after around 3 minutes and the device then switches off automatically. 
 
Fever alarm 
If the temperature measured is 37.5°C or higher, the fever alarm beeps (short double beeps for 
about 4 seconds) and the LC display flashes.  
 
Display of last measured value 
If you keep the I/O button pressed for longer than three seconds, the last measured value will 
be shown in the display. As soon as you release the I/O button, the thermometer returns to 
measurement mode.
 
 
Switches the display from °C to ° F and from °F to °C, resp. 
Centigrade or Fahrenheit mode can be switched within the first 3 seconds by pressing and  
holding the I/0 button. 
Measured value deviation 
The measured temperatures will always deviate, depending on the measurement location. This  
is normal, and caused by deviations in temperature distribution on the human body. You could,  
for  example,  measure  a  rectal  temperature  of  36.5°C,  an  oral  temperature  of  36.0°C  and  an 
underarm value of 35.5°C. 
  
 
Measurement locations 
In the mouth (oral) 
Small 
temperature 
variations 
in 
the 
mouth 
between 
measurements  are  normal.  These  are  caused  by  varying 
temperature  distribution  within  the  mouth.  Try  to  place  the 
thermometer  in  one  of  the  two  heat  pockets under  the  tongue, 
to  the  left  and  right  of  the  tongue  root.  The  measurement 
sensor  must  have  good  contact  with  the  tissue.  The  subject 
should close the mouth and breathe normally through the nose 
so  that  the  measured  result  is  not  affected  by  breathed  air. 
Measuring period: approx. 10 seconds. 
 
In the anus (rectal) 
This is the most accurate method. It is particularly suitable for infants and small children. 
Carefully guide the tip of the thermometer approximately 1  cm into the anus. Measuring period: 
approx. 10 seconds. 
 
Under the arm (axillary) 
From a  medical viewpoint this type of measurement leads to inaccurate values. It is 
therefore unsuitable for carrying out exact measurements. 
 
 
Changing the battery 
As  soon  as  the 
  symbol  appears  in  the  display,  the  battery  has  run  out  and  requires 
replacement. Always replace the discharged battery with one of the same type (LR 41 or AG3). 
To do this, proceed as follows: 
Unlock  the  screw  at  the  rear  side  of  the  thermometer.  Pull  the  transparent  part  from  the 
thermometer. Now pull carefull the flexible measuring point from the thermometer body.  
 
 
 
 
 
 
 
 
ATTENTION: Remove the flexible measuring point not longer as 2 cm from the  
                       thermometer body. Otherwise you damage the connection wires.  
 
Remove the used battery with the help of a pointed object (pincers, small screwdriver etc.) and 
put the new battery in. Please make sure it is the right way round. You can purchase batteries 
from jewellers or electrical shops. 
 
Measurement checks 
Checking  the  accuracy  of  the  SC  29  flex  is  not  necessary  in  most  cases.  If  you  require  a 
measurement check (at your expense), please send the SC 29 flex to our customer service  
address. Please make sure it is well-packaged and that the correct postage stamps are applied. 
Cleaning and disinfection 
Please only use an alcohol pad. Alcohol pads can be purchased from any dispensing chemist.  
We  recommend  the  use  of  Mucocit
-F2000  or  Helipur
Hplus  to  disinfect  the  thermometer 
when  it  is  being  used  in  medical  facilities.  Both  are  effective  against  all  hospital-typical 
pathogens and can be purchased from any dispensing chemist. 
 
Guarantee 
We give a guarantee period of 2 years from the date of purchase. We guarantee perfect design 
of this article. We will rectify all material or manufacturing errors during the guarantee period. If, 
contrary  to  expectations,  defects  are  found,  please  send  the  device  back  to  our  customer 
service  department.  Please  enclose  the  proof  of  purchase  and  a  description  of  the  defect  so 
that  the  problem  can  be  rectified  as  soon  as  possible.  The  guarantee  excludes  damages 
caused by improper handling, related to the battery or to the packaging. 
 
Disposal note                                                                                                 
The  disposal  of  batteries  is  regulated.  Spent  batteries  are  collected  at  collection  points 
provided by battery dealers and the communities. Dispose
 
of
 
the
 
thermometer
 
monitor
 
as 
prescribed. i.e.,
 
at
 
a collection point for 
electrical scrap. Dispose of the thermometer monitor as prescribed. i.e., at a collection point for 
electrical scrap. 
 
Specifications 
Type: 
maximum thermometer 
Measuring range: 
32.0°C to 43.9°C 
 
Temperature below 32.0°C: display "L
“ for low. 
 
Temperature above 43.9°C: display "H“ for high. 
Measurement accuracy: 
0.1°C between 35.5°C and 42.0°C (0.2°C other range) at a 
 
 
surrounding temperature of 18°C to 28°C. 
Display: 
liquid-crystal (LCD display). 
Battery type: 
1 no.: LR 41 or AG3. 
Battery display: 
The symbol 
 shows when the battery is discharged. 
Storage temperature: 
0°C to 50°C 
 
Referenced norms 
The  device  has  been  certified  in  accordance  with  DIRECTIVE  93/42/EU  of  14th  June  1993 
relating  to  medicinal  products.  The  clinical  thermometer  SC  29  flex  complies  with  the 
requirements of European norm EN 12470-3:2000.  
  
Explanation of symbols  
 
 
/ service client 
 
 
 
 
 
SCALA Electronic GmbH  
Ruhlsdorfer Str. 95 
14532 Stahnsdorf 
Germany 
www.scala-electronic.com 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
Rev. 06/05/2014/GB,F