Beurer ME80 ユーザーズマニュアル

ページ / 52
52
Table 4
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF communications equipment and the 
ME EQUIPMENT or ME SYSTEM
The ME80 is intended for use in an electromagnetic environment in which radiated RF disturbances are con-
trolled. The customer or the user of the ME80 can help prevent electromagnetic interference by maintaining a 
minimum distance between portable and mobile RF communications equipment (transmitters) and the ME80 as 
recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
(W)
Separation distance according to frequency of transmitter
(m)
150 kHz to 80 MHz
d = 1.2 
80 MHz to 800 MHz
d = 1.2 
800 MHz to 2.5 GHz
d = 2.3 
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in 
meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the 
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1  At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2   These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorp-
tion and reflection from structures, objects and people.
752.976 - 0114_ Irr
tum und Änderungen vorbehalt
en