Valco electronic Valve Oven ユーザーズマニュアル

ページ / 62
Chapter 4
Troubleshooting
44
Refer to the TRACE operations manual, Chapter 4, Gases and Controls, for leak 
test suggestions and the necessary hardware required to pressure check a 
plumbing system. 
Valve Switching
Valco valves contain a rotor that is turned by the actuator. This rotor channels the 
gas between the various ports of the valve. If the rotor is not being switched 
properly, a flow path may be partially or fully blocked, or leaks may occur 
between valve ports. 
Each valve contains a mechanical stop that should reach an extreme when the 
valve is rotated in the clockwise or counterclockwise direction. Close inspection 
of the valve will indicate if the valve is being switched to its limits, Figure 2-4. 
Actuator pressure should be 420 Kpa (60psig) to 630 Kpa (90 psig). 
Table 4-1. Leak Testing for Valves
Symptom
Cause
Remedy
Valve leaking between ports
Defective rotor
Replace rotor or valve
Valve leaking around rotor
Valve rotor not seating 
properly
Rotor Defective
Tighten rotor
Replace Rotor in valve
Leaking out valve ports
Ferrule not seating
Replace ferrule
Table 4-2. Causes and Remedies for Improper Valve Switching
Cause
Remedy
Low actuator line pressure
Increase line pressure
Solenoid not actuating
Verify timed event is set correctly
Check voltage at solenoid 
Replace defective solenoid