SETUP Menu
Teledyne API - T100 UV Fluorescence SO2 Analyzer
100
To enable or disable passwords, press:
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SAMPLE RANGE = 500.0 PPB SO2 =XXX.X
< TST TST > CAL
SETUP
SETUP
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS
CLK MORE
EXIT
SETUP
PASSWORD ENABLE: OFF
OFF
ENTR EXIT
SETUP
PASSWORD ENABLE:
ON
ON
ENTR EXIT
SETUP
PASSWORD ENABLE: ON
ON
ENTR
EXIT
EXIT
ignores the
change.
ENTR
accepts the
change.
Once Password is
enabled, exit back
enabled, exit back
out to the main
menu for this feature
to take effect.
Toggles
password
status On/Off
ENTR
accepts
displayed
password value
EXIT
returns to
SAMPLE
display
PASSWORD
default state is
OFF
Figure 5-9: SETUP – Enable Password Security
If the password feature is enabled, then when entering either Calibration or Setup Mode,
the default password displayed will be 000, and the new password must be input.
Example follows for Calibration Mode:
the default password displayed will be 000, and the new password must be input.
Example follows for Calibration Mode:
06807C DCN6650