Data Acquisition System (DAS) and APICOM 
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
168 
7.2.7. RS-232 REPORT FUNCTION 
The M DAS can automatically report data to the communications ports, where they can 
be captured with a terminal emulation program or simply viewed by the user.  
To enable automatic COMM port reporting, follow the instruction shown in Figure 7-3, 
then press: 
 
 
 
 
 
Edit Data Channel Menu
 
 
 
SETUP X.X     
  0) CONC:     ATIMER,     2,       4032, R
 
 
 PREV  NEXT            INS    DEL
   EDIT    
 PRNT 
   EXIT 
Exits to the main 
Data Acquisition 
menu 
 
 
 
SETUP X.X      
NAME:CONC
 
 
    <SET  
  SET> 
 EDIT
 
   PRINT 
 
EXIT
 
 
 
 
SETUP X.X      
RS-232 REPORT: OFF
 
 
    <SET    SET>
  EDIT 
   PRINT 
 
EXIT
 
ENTR
 accepts the new 
setting and returns to the 
previous menu. 
EXIT
 ignores the new setting 
and returns to the previous 
menu. 
SETUP X.X        
RS-232 REPORT: OFF
 
  
 OFF   
    
 
 
 
    ENTR    EXIT
 
 
Toggle to turn 
reporting ON or OFF 
Press SET>  until… 
 
 
Figure 7-10: 
DAS – RS-232 Report Function 
7.2.8. COMPACT REPORT 
When enabled, this option avoids unnecessary line breaks on all RS-232 reports. Instead 
of reporting each parameter in one channel on a separate line, up to five parameters are 
reported in one line, instead. 
7.2.9. STARTING DATE 
This option allows a user to specify a starting date for any given channel in case the user 
wants to start data acquisition only after a certain time and date. If the Starting Date is 
in the past, the DAS ignores this setting. 
7.2.10. DISABLING/ENABLING DATA CHANNELS 
Data channels can be temporarily disabled, which can reduce the read/write wear on the 
disk-on-module. The ALL_01 channel of the T100, for example, is disabled by default.  