Teledyne API - T100 UV Fluorescence SO2 Analyzer
Calibration Procedures
197
STEP TWO: Perform the zero/span check.
Wait until STABIL
falls below 0.5
ppb. This may
ppb. This may
take several
minutes.
The value of STABIL
may jump
significantly. Wait
until STABIL falls
until STABIL falls
below 0.5 ppb. This
may take several
minutes.
SAMPLE
STABIL=XXX.X PPB
SO2 =XXX.X
< TST TST > CAL
CALZ
CALS
SETUP
ZERO CAL M
STABIL=XXX.X PPB
SO2 =XXX.X
< TST TST >
ZERO
CONC
EXIT
SAMPLE
STABIL=XXX.X PPB
SO2 =XXX.X
< TST TST > CAL
CALZ
CALS
SETUP
EXIT
returns to the main
SAMPLE
display
ACTION:
Record the
SO
2
readings
presented in the
upper right corner of
the display.
ACTION:
Record the
SO
2
readings
presented in the
upper right corner of
the display.
Scroll to the STABIL test
function.
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST >
CAL
CALZ
CALS
SETUP
SPAN CAL M
STABIL=XXX.X PPB
SO2 =XXX.X
< TST TST >
ZERO
SPAN
CONC
EXIT
Figure 9-10:
Setup for Manual Calibration Check with Z/S Valve or IZS Option (Step 2)
06807C DCN6650