Teledyne API - T100 UV Fluorescence SO2 Analyzer  
92 
5.4.3. REPORTING RANGE MODES: SINGLE, DUAL, AUTO RANGES 
The T100 provides three analog output range modes to choose from:  
  Single range (SNGL) mode sets a single maximum range for the analog output. If 
single range is selected (refer to Section 5.4.3.1) both outputs are slaved together 
and will represent the same measurement span (e.g. 0-50 ppm), however their 
electronic signal levels may be configured for different ranges (e.g. 0-10 VDC vs. 0-
.1 VDC – Refer to Section 5.9.3.1). 
  Dual range (DUAL) allows the A1 and A2 outputs to be configured with different 
measurement spans (refer to Section 5.4.3.2). 
  Auto range (AUTO) mode gives the analyzer to ability to output data via a low 
range and high range. When this mode is selected (refer to Section 5.4.3.3) the T100 
will automatically switch between the two ranges dynamically as the concentration 
value fluctuates.  
Also, in this mode the RANGE Test function displayed on the front panel during 
SAMPLE mode will be replaced by two separate functions, Range1 and Range2.   
 
Range status is also output via the External Digital I/O Status Bits (refer to Section 
8.1.1). 
To select the Analog Output Range Type press: 
 
 
 SAMPLE          RANGE = 500.000 PPB        SO2 =XXX.X   
 
 < TST  TST >  CAL  
 
 
SETUP
 
SAMPLE          
ENTER SETUP PASS : 818
 
 
        
 8        1        8 
 
                      
ENTR
  
 EXIT
 
 
 
 
SETUP X.X 
 
 CFG  DAS
  RNGE  
PASS  CLK  MORE  
EXIT
 
 
 
SETUP X.X
RANGE CONTROL MENU
 
 
 
MODE
 
 
 
 
SET  UNIT
 
 
 
 
EXIT
 
 
 
SETUP X.X
RANGE MODE:
SNGL
 
 
 SNGL
 
 
 
 DUAL  AUTO
 
 
     ENTR  EXIT
 
EXIT
Returns
to the Main 
SAMPLE
 Display 
Go To 
Section 
6.7.4 
Go To 
Section 
6.7.5 
Go To 
Section 
6.7.6 
Only one of the 
range modes may 
be active at any 
time. 
 
 
Figure 5-3: SETUP RNGE – Reporting Range Mode 
 