APPENDIX A Firmware Commands 
Teledyne API Ultrafine Particle Monitor - Model 651 
A-6 
07506C DCN6727 
RIE – Read Instrument Errors 
RIE reads the instrument errors (displayed as a 16-bit integer in 
hexadecimal format). The number may be a combination of the 
values of more than one flag. 
RIE 
Command RIE 
 
Response 
Bit 0 x XXXX  Bit = Hexadecimal character A-F. When the bit 
is set, the parameter is in error. 
XXXX = 4-digit number 
0x0001 = Conditioner Temperature 
0x0002 = Growth Tube Temperature 
0x0004 = Optics Temperature 
0x0008 = Vacuum Level 
0x0020 = Laser Status 
0x0040 = Water Level  
0x0080 = Concentration Over-range 
0x0100 = Pulse Height Fault 
0x0200 = Absolute Pressure 
0x0400 = Nozzle Pressure 
0x0800 = Water Separator Temperature 
0x1000 = Warmup 
0x2000 = Reserved 
0x4000 = Service Reminder 
0x8000 = Reserved 
Example
 
Command RIE 
Read 
Instrument Errors 
Response C00 
Water Separator Temperature and Nozzle 
Pressure faults (Nozzle Pressure = 
hexadecimal 4. Water Separator = 
hexadecimal 8. Added together they make 
hexadecimal C.)  
Note:  Hexadecimal is a numerical system using a base of 16. The 
symbols 0-9 represent the values zero to nine, and the letters 
A-F represent the values ten to sixteen. It is a useful 
“shorthand” for computer engineering because each 
hexadecimal digit represents four binary digits.  
RIF – Read Aerosol Flow Rate 
RIF reads the inlet flow rate in liters per minute (L/min). 
RIF 
Command RIF 
 
Response X 
X = Floating point number either 0.12, 0.6, or 
3.0 
Example
 
Command 
RIF 
Read Inlet Flow Rate 
Response 0.3 
3.0 
L/min