Teledyne Computer Monitor 651 ユーザーズマニュアル

ページ / 127
Maintenance, Service, and Troubleshooting 
Teledyne API Ultrafine Particle Monitor - Model 651 
86 
07506C DCN6727 
Note the orientation of the orifice. 
Orifice Insert
 
Figure 9-10 
Orifice Insert  
3.  Grasp the tubing on either side of the orifice and pull firmly to 
detach the tubing from the orifice.  
4.  Using a microscope, inspect the orifice. If debris is present, soak 
the orifice in isopropyl alcohol for 20 minutes. 
Inspect the Orifice
 
Figure 9-11 
Inspect Orifice 
5.  Using compressed air at <60 psi, blow out the orifice and then 
re-inspect under the microscope. 
6.  If the orifice is clean, replace it in the instrument making sure 
that the orifice insert is positioned closest to the manifold block. 
7.  If the orifice is not clean, replace it with a new one. 
Note: replacement orifices are available from TAPI  
8.  Replace the instrument cover.  
I n s p e c t   L i q u i d   L i n e s  
Inspect the water filling lines that flow from the fill connectors 
located on the front and back panels to the fill valve. Also inspect 
the water line from the fill valve to the Model 651 engine. Check for 
cracks, damage, loose fit, or signs of leaking. Replace as necessary 
or annually, whichever comes first, with tubing supplied in the 
maintenance kit that comes with the instrument.