5 • Service
Doc. 065291-01 3/09
107
3. On the Chromeleon Wellness Panel under Pump Flow Rate 
Calibration, select Reset Cal, or touch 
Reset Cal
 on the touch 
screen.
4. Select 
1 mL/min
 to set the flow rate and begin pumping deionized 
water.
5. Allow the pump to stabilize for 20 minutes.
6. If you are using the Wellness Panel, click 
5.00 min 
to start the timer. 
If you are using the touch screen, start an external timer. Immediately 
start collecting water into the tared beaker. 
7. Collect the water for exactly 5.00 minutes.
8. Enter the weight of the water (in grams).
9. If you are using the Wellness Panel, click 
Calibrate
 to download the 
value to the ICS-2100.
10. To recheck the value, wait at least 15 minutes, and then repeat the 
calibration. 
11. If you are using the Wellness panel, click 
Log
 to record the new 
calibration value in the Audit Trail. 
12. If you performed the calibration from the ICS-2100 touch screen, 
record the new calibration values in the Chromeleon Wellness 
database. To do this, open the Wellness Panel (see 
click Upload.