© Copyright 1970-2008 Kirby Morgan Dive Systems, Inc. All rights reserved. Document # 080508001
Kirby Morgan 37 & 57
Chapter 7.0
Breathing System Maintenance and Repairs
7.1 Introduction
This chapter covers the maintenance and repair of all 
components of the breathing system. The breathing 
system  includes  the  one  way  valve,  the  emergency 
valve,  the  side  block,  the  bent  tube  assembly,  the 
demand regulator, and the oral/nasal mask. 
The breathing systems on all Kirby Morgan helmets 
and masks are simple and highly reliable. The fact that 
they can continue to operate when the components are 
not in a well-maintained condition can cause divers 
to become complacent about maintenance. 
Your life depends on the correct function of this 
equipment!
While Kirby Morgan helmets and masks are simple 
to maintain, like any type of life support equipment, 
they do require regular periodic maintenance to func-
tion properly.
All parts disassembled should be thoroughly cleaned 
using the methods described in the Appendix in the 
rear of this manual. Components that require the use 
of lubricants, sealing and thread locking compounds 
should also be maintained. 
Most fasteners have a torque value, it is imperative that 
all fasteners which have a torque value be tightened 
to the torque specifications as outlined by the proce-
dure, or as listed in appendix 1. If in doubt as to  the 
proper torque setting, contact your local authorized 
repair facility or KMDSI.
7.2 One Way Valve
NOTE: The one-way valve assembly should be disas-
sembled,  cleaned  and  the  three  O-rings  should  be 
replaced at least annually. Damaged and/or corroded 
parts should be replaced. A repair kit is available for 
replacement parts (525-330).
7.2.1 Disassembly Of The One Way Valve
Tools Required:
Soft Jaw Vice
1  inch  Open  End  Wrench  Attachment  on  Torque 
Wrench
(If no vise is available use a backup 1 inch open end 
wrench)
To  disassemble  and  inspect  the  one  way  valve  as-
sembly:
1)  The one way valve assembly must be removed 
from  the  side  block.    Use  the  open  end  wrench  to 
remove it.
2)  After the one way valve has been removed, use two 
wrenches or hold the hex part of the body in a soft jaw 
vise while removing the seat with a wrench.
 
As the seat is removed, the wiper  and the O-ring  
will slide out in place in a groove on the seat. The 
Use two wrenches or hold the hex part 
of the body in a vise while removing or 
turning  the  seat  with  a  wrench.  Do  not 
use pliers on the main body of the one-
way valve. You may damage the valve if 
pliers are used.
CAUTION