Specifications
51-52-25-150, Issue 1 – April 2009
Conditions For Use
Reference Test Conditions
Ambient Temperature:
20C ±2C.
Relative Humidity:
60 to 70%.
Supply Voltage:
100 to 240V AC 50Hz ±1%.
Source Resistance:
<10Ω for thermocouple input.
RTD Lead Resistance:
<0.1Ω/lead balanced (Pt100).
Operating Conditions
Ambient Temperatures 
0C to 55C (operating) and -20C to 80C (storage).
Relative Humidity:
20% to 95% non-condensing.
Altitude:
Up to 2000m above sea level.
Supply Voltage:
Either 100 to 240V ±10% AC 50/60Hz
or 20 to 48V AC 50/60Hz & 22 to 55V DC for low voltage versions.
Power Consumption:
Mains versions:              24VA.
Low voltage versions:    15VA / 12W.
Source Resistance:
1000Ω maximum (thermocouple).
RTD Input Lead 
Resistance:
50Ω per lead maximum, balanced
Standards
Conformance Norms:
CE, UL, ULC.
EMC standards:
Complies with CE EN61326.
Safety Standards:
Complies with CE EN61010-1 and UL3121.
Pollution Degree 2, Installation Category II.
Front Panel Sealing:
To IP66 (IP65 front USB connector). IP20 behind the panel.
(IP ratings are not tested for or approved by UL)
Dimensions
Front Bezel Size:
1
/
4
DIN (96 x 96mm).
Mounting:
Plug-in with panel mounting fixing strap.
Panel Cut-out Size:
92mm x 92mm. Max panel thickness 6.0mm
Depth Behind Panel:
117mm
Ventilation
20mm gap required above, below and behind.
Weight:
0.65kg maximum.
Terminals:
Screw type (combination head).